Code
Error Title
Error Cause Text
Error Message
200
Faulty Delivery
Device
Unable to Read from
Delivery Device Memory
Remove and reconnect Delivery Device electrical cable. If problem persists,
replace Delivery Device.
205
Faulty Delivery
Device
Unable to Write to
Delivery Device Memory
Remove and reconnect Delivery Device electrical cable. If problem persists,
replace Delivery Device.
210
Faulty Delivery
Device
Faulty Delivery Device
Thermocouple
Replace Delivery Device.
211
Faulty Delivery
Device
Faulty Delivery Device
Trigger Signals
Replace Delivery Device.
215
Faulty Delivery
Device
Invalid Therapy Code
Remove and reconnect Delivery Device electrical cable. If problem persists,
replace Delivery Device.
218
Faulty Delivery
Device
Delivery Device
Impedance Error
Replace Delivery Device.
219
Faulty Delivery
Device
Delivery Device Frequency
Error
Replace Delivery Device.
220
Expired Delivery
Device
Maximum Full Treatments
Exceeded
Replace Delivery Device.
225
Faulty Delivery
Device
Delivery Device is
Permanently Disabled
Replace Delivery Device.
230
Expired Delivery
Device
Maximum Vapor Time
Exceeded
Replace Delivery Device.
235
Prime Failed
Low Temperature (Prime) Replace Delivery Device.
236
Pre-Treatment Failed Low Temperature (Pre-
Treatment)
Replace Delivery Device.
240
Prime Failed
Low Water Pressure
(Prime)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace
syringe and reprime Delivery Device. If leaks are found, replace Delivery
Device. If no bubbles or leaks are observed, replace Delivery Device.
241
Prime Failed
High Water Pressure
(Prime)
Check water line for kinks. Resume priming. If problem persists, replace
Delivery Device.
242
Pre-Treatment Failed Low Water Pressure (Pre-
Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace
syringe and reprime Delivery Device. If leaks are found, replace Delivery
Device. If no bubbles or leaks are observed, replace Delivery Device.
243
Pre-Treatment Failed High Water Pressure (Pre-
Treatment)
Check water line for kinks. Resume pre-treatment vapor cycle. If problem
persists, replace Delivery Device.
245
Prime Failed
High Temperature (Prime) Check syringe and water line for bubbles or leaks. If bubbles are found, replace
syringe and reprime Delivery Device. If leaks are found, replace Delivery
Device. If no bubbles or leaks are observed, replace Delivery Device.
246
Pre-Treatment Failed High Temperature (Pre-
Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace
syringe and reprime Delivery Device. If leaks are found, replace Delivery
Device. If no bubbles or leaks are observed, replace Delivery Device.
250
Prime Paused
Trigger Released
Delivery Device trigger was released before priming completed. Press and hold
Vapor Activation Button (blue) to continue priming.
251
Pre-Treatment
Paused
Trigger Released
Delivery Device trigger was released before pre-treatment vapor cycle
completed. Press and hold Vapor Activation Button (blue) to restart pre-
treatment vapor cycle.
255
Treatment Halted
Low Temperature
(Treatment)
Retract the needle and remove the Delivery Device from patient. Replace
Delivery Device.
260
Treatment Halted
High Water Pressure
(Treatment)
Check water line for kinks. Resume treatment. If problem persists, replace
Delivery Device.
265
Treatment Halted
Low Water Pressure
(Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace
syringe and reprime Delivery Device. If leaks are found, replace Delivery
Device. If no bubbles or leaks are observed, replace Delivery Device.
270
Syringe is Empty
Syringe Empty
Retract the needle and remove the Delivery Device from patient. Replace
syringe and reprime Delivery Device.