3.1.3
Priming the Delivery Device
CAUTION: When any of the following occur, turn the power off, remove all accessories from the
patient, and unplug the power cord from the outlet.
There is smoke or a strange odor leaking out of the generator.
The generator has been dropped or impacted by an object.
Liquid or foreign matter gets inside the generator.
If you think the generator may have been damaged.
Figure 11: Ready for Priming
When the operator starts priming the Delivery Device, a progress bar will be initiated and
displayed. The text will change, and a tone will be generated to indicate that the operation is in
progress. Priming is initiated by deploying the needle and holding the Delivery Device vapor
activation button for the priming duration, which lasts approximately 30 seconds. A message
will be displayed indicating when Priming is complete.
If the vapor activation trigger is activated while the needle is being deployed, priming, pre-
treatment, or treatment operation will not be initiated until the vapor trigger is released and
reengaged.