3033-001 Revision D January 2018
©NxThera 2018
Page 11 of 90
1.5.1.1
DANGERS
DANGER:
Do not take or use the device in locations where combustible anesthetics or
flammable gases are used or in high-pressure oxygen rooms or inside oxygen
tents.
1.5.1.2
WARNINGS
WARNING:
A protective ground connection by way of the grounding conductor in the power
cord is essential for safe operation. To avoid electrical shock, plug the power cord
into a properly wired receptacle, use only the power cord supplied with the
generator, and make sure the power cord is in good condition.
WARNING:
After visual inspection, if the generator is damaged or a message is indicated to
not use the generator, please contact Customer Service and take the generator out
of service.
WARNING:
Before conducting maintenance work, turn the power off and unplug the power
cord from the outlet to prevent electric shock.
WARNING:
Do not modify this equipment without authorization of NxThera.
WARNING:
Carefully read and understand all instructions, indications, warnings, and cautions
in this operator’s manual prior to using the Rezūm Generator. Failure to do so
could result in compromised patient safety, patient complications and/or
insufficient treatment.
WARNING:
Do not connect a grounding wire from a grounding stud to a gas pipe or water
pipe.
WARNING:
Do not connect to an electrical outlet controlled by a wall switch because the
device may be accidentally turned off.
WARNING:
Do not plug power cord into an outlet (or unplug it) with wet hands.
WARNING:
Do not submerge the device in liquids or pour cleaning liquids over, into or onto
the generator.
WARNING:
Do not use the generator if it is damaged, is not functioning properly, or fails to
meet an electrical safety check. Notify the appropriate personnel to ensure the
generator is removed from service and properly repaired.