Revitive Medic 5573AQ User Manual Download Page 11

20

21

j   

If you remove one or both feet from the foot-pads, Revitive will pause the  

 

 

 

stimulation for up to 10 minutes. The display will blink on and off and you  

 

 

 

will hear a beep. The timer will stop until the session is resumed. Do not  

 

 

 

touch the foot-pads with your hands or pick up device until paused is enabled

 

 

The stimulation can be resumed by placing both feet back on the foot-pads and  

 

 

then pressing the        (+) button on the Intensity Control.  

 

 

 

While Revitive is paused, the intensity level can be reduced by pressing the          

 

                (–) button on the Intensity Control.    

 

 

Intensity cannot be increased while Revitive is paused.  

 

 

 

After 10 minutes of being in the paused state, Revitive will switch itself off.

k

  

The        Time Display will start to count down in minutes as Revitive cycles  

 

 

 

through its program. 

l   

When Revitive times out, the        Time Display reads 00 and you will hear  

 

 

 

three beeps. 

m

  

Revitive switches itself off automatically.

Instructions for use

g  Press the        (+) Intensity Control one time to  

 

 

start stimulation. 

 

You can use the buttons on the control panel or  

 

 

remote control. Ensure both feet are on the foot-   

 

pads when increasing the intensity.  

 

 

Press the (+) button to increase the level. 

 

Press the (–) button to decrease the level.

70
3o

h

 

Ensure you familiarise yourself with the feeling of the Medic Program. Use on a low   

 

intensity for 30-60 minutes once per day for a few days.

 

Set the intensity at a level where you can feel the stimulation in your feet which  

 

 

creates a calf muscle contraction. 

 

It is normal to feel tingling or varying sensations in your feet and calf muscles.

 

Revitive does 

not vibrate. 

i  After a few days of familiarisation, use a higher intensity that provides you with  

 

 

strong but comfortable calf muscle contractions.

 

Most people experience a strong contraction at intensity level 40 or over.    

 

The IsoRocker® enables Revitive to rock up and down once muscle contractions are  

 

nearing the desired level. The rocking is caused by muscle contractions in the leg,    

 

and the resulting ankle movement allowed by the IsoRocker is a good indicator  

 

 

that the intensity level is sufficient to provide the therapy. Increasing the intensity    

 

above this level will enhance the therapy, but should not be allowed to become    

 uncomfortable. 
 

You can set the intensity between 1-99. 

 

Please note that the intensity required may vary from day to day.    

If you feel pain or discomfort:    

 

 

 

 

•    Remove one or both feet from the foot-pads to stop the stimulation  
 

•   Lower the intensity of the foot-pads using  the        (–) Intensity Control on    
 

the  control panel, or remote control, before replacing your feet on Revitive

5603_IFU02_17807098.indd   20-21

5603_IFU02_17807098.indd   20-21

12/01/2021   08:53

12/01/2021   08:53

Summary of Contents for Medic 5573AQ

Page 1: ...this product USE ONLY AS DIRECTED If symptoms persist consult your healthcare professional Medic Drug Free Clinically Proven Certified Medical Device User s Manual CIRCULATION BOOSTER 5603_IFU02_1780...

Page 2: ...21 Step 4 Using the Dual Mode Body Pads 22 27 4 1 Switching between EMS and TENS mode 27 Step 5 Cleaning and storing 28 Replacing the batteries Body Pads 28 Troubleshooting 29 31 Technical Specificat...

Page 3: ...his booklet Power adapter Cable length 180cm Revitive Medic Electrode Body Pads Electrode cord Cable length 150cm Remote control AAA batteries x 2 1 2 3 4 5 6 Parts and controls 70 3o Carry handle Con...

Page 4: ...pads Revitive is intended to Reduce pain and discomfort in the legs ankles feet caused by poor circulation Peripheral Arterial Disease Improve circulation in the legs caused by Peripheral Arterial Dis...

Page 5: ...tor as soon as possible if you show any signs of the above symptoms Important safeguards Introduction to Revitive Electrical Muscle Stimulation EMS Indications for use continued When using electrode p...

Page 6: ...tenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Revitive System including cables specified by the manufacturer Important safeguards Warnings Consult...

Page 7: ...operating temperatures to ensure continued product performance Cautions When applying foot pads or electrode body pads Be careful when applying stimulation over areas of skin that lack normal sensatio...

Page 8: ...elp identify the parts b Locate the remote control and insert the batteries Before you use Revitive for the first time read the Important Safeguards on pages 9 14 If in doubt consult your doctor befor...

Page 9: ...ry level If you turn the device off and the low battery level indicator blinks for 10 seconds the battery requires recharging Please note you can use the device while it is charging but the battery in...

Page 10: ...aveforms to your feet and legs You can use the Body Pads to receive EMS or TENS see pages 20 25 for details a It is important that you are well hydrated We recommend that you apply a moisturiser to th...

Page 11: ...arise yourself with the feeling of the Medic Program Use on a low intensity for 30 60 minutes once per day for a few days Set the intensity at a level where you can feel the stimulation in your feet w...

Page 12: ...provides you with strong but comfortable muscle contractions It is recommended to use Revitive in TENS mode for temporary pain relief from chronic joint muscular nerve or arthritic pain Use for 20 30...

Page 13: ...se the default time of 30 minutes set the time for the number of minutes you want You can use the controls on the control panel or on the remote control Press the Time Control to increase the time Pre...

Page 14: ...imulation You can set the intensity between 1 99 You can use the buttons on the device or remote control Press the button to increase the level Press the button to decrease the level It is normal to f...

Page 15: ...lectrode cord is plugged in Revitive may be in auto pause Ensure that you have drunk plenty of water and that your feet are well moisturised Place both feet onto the foot pads then press intensity but...

Page 16: ...rrectly If still not working contact your authorised dealer Flat battery Plug device in an electrical socket and charge battery or run using mains power option Revitive is not vibrating REVITIVE IS NO...

Page 17: ...e Rectangular Maximum Output Voltage 500 50V Pulse Duration 100 s Net Charge per pulse 500 0 C Maximum Average Power Density 500 4 19 mW cm2 Technical Specifications Name of product Revitive Medic Mod...

Page 18: ...C allow the product to re adjust to the recommended operating temperatures to ensure continued product performance Indoor Use Only Medical device does not contain natural rubber latex Do not disassem...

Page 19: ...to retain your original till receipt or invoice to indicate the date of initial purchase To qualify for the 2 year guarantee the device must have been used according to the manufacturer s instructions...