Controlling the device using the touchscreen display
50
Fig. 33:
Display with connected DR 100 after closing the lid
Element Description
Function
T1
Status display
Automatic lid opening switched on
– Acoustic signal on/off.
T2.1
System settings
Access to the system settings.
T2.2
Program settings
Access to program settings.
T3.1
Speed
Displays the set speed (6,000 - 23,000 rpm).
T3.2
Start delay
Device start in minutes/seconds (mm:ss).
T3.3
DR 100 conveying speed
Setting the conveying speed of the DR 100 feed unit. This
function can only be selected when the DR 100 feed unit is
connected and switched on. The conveying speed set
manually on the feed unit is overwritten by the value set on
the device when the grinding process is started at the device.
T3.4
Automatic stop
Automatic stop of the device after mm:ss.
The device detects when there is no grinding load and
automatically stops the grinding process and the sample
material feed after the specified time. This function can only
be selected when the DR 100 feed unit is connected and
switched on. This function is only active when the grinding
process is started via the device.
T4.1
Open
Opens the lock of the device hood.
T4.2
Start
Starts the grinding process.
T3.3
T3.4