6-2
REM
staR
M s
ERiEs
U
sER
M
anUal
6.2 a
dding
s
UpplEMEntal
o
xygEn
Oxygen may be added at the mask connection. Please note the warnings listed below when using
oxygen with the device.
W
aRnings
• When using oxygen with this system, the oxygen supply must comply with local regulations
for medical oxygen.
• When using oxygen with this system, a Respironics Pressure Valve (Part number 302418)
must be placed in-line with the patient circuit. Failure to use the pressure valve could result
in a fire hazard.
• Oxygen accelerates fires. Keep the device and the oxygen container away from heat, open
flames, any oily substance, or other sources of ignition. Do not smoke in the area near the
device or the oxygen.
• When using oxygen with this system, turn the device on before turning on the oxygen. Turn
the oxygen off before turning the device off. This will prevent oxygen accumulation in the
device.
• If administering fixed-flow supplemental oxygen, the oxygen concentration may not be con-
stant. The inspired oxygen concentration will vary, depending on the CPAP setting, patient
breathing pattern, and leak rate. Substantial leaks around the mask may reduce the inspired
oxygen concentration to less than the expected concentrations. Appropriate patient monitor-
ing should be implemented.
Summary of Contents for REMstar M Series
Page 1: ...1016452 AM 2 09 06 ...
Page 22: ...2 4 REMstar M Series User Manual ...
Page 30: ...3 8 REMstar M Series User Manual ...