The device does not
operate when you
press the button.
The airflow does not
turn on.
There may be a prob-
lem with the blower.
Make sure the device is powered correctly,
following the instructions on the previous
page. If the button backlights turn on when
you apply power, but the airflow does not
turn on, there may be a problem with your
device. Contact your home care provider for
assistance.
Note:
When the device is functioning cor-
rectly, after you press the Start/Stop
button
, the airflow turns on
after a slight delay. This brief delay is
normal.
The device’s display is
erratic.
The device or power
supply has been
dropped or mishan-
dled, or the device or
power supply is in an
area with high Electro-
magnetic Interference
(EMI) emissions.
Unplug the device and the power supply.
Reapply power to the device. If the problem
continues, relocate the device to an area with
lower EMI emissions (e.g., away from elec-
tronic equipment such as cellular phones,
cordless phones, computers, TVs, electronic
games, hair dryers, etc.).
If the problem still occurs, contact your home
care provider for assistance.
Device Resets/
Reboots
The device shuts down
and restarts automati-
cally during therapy.
(This is unlikely to
occur.)
The device comes
installed with trouble-
shooting software that
automatically monitors
performance.
Such a reset poses no danger to the pa-
tient and assures that the patient receives
prescribed therapy throughout the night. If
there is a possibility of damage to the device,
the device will shut down permanently.
The product will then display the following
system alert symbol
to indicate that the
device must be returned to the home care
provider for service.