I
mportant
! Hand wash mask before use.
Before adjusting the mask and headgear,
unhook the EZPeel™ tabs on the Premium
Headgear and remove it from the mask. Clean
the mask following the instructions below.
C
leanIng
I
nstrUCtIons
For Multi-Patient Use Reusable. Contact Res-
pironics Customer Service at 1-724-387-4000 or
+49 (0) 8152 93060 to obtain Mask Disinfecting
instructions (Order #1011138).
For Home Use: Begin by removing or discon-
necting the headgear from the mask. Clean
the headgear at least once each week or more
often if needed. Fasten the hook tabs on the
straps so that they don’t catch other garments in the laundry.
m
ask
and
C
UsHIon
Thoroughly clean the mask each day, immediately after use, using the hand
wash instructions. In addition to hand washing, the mask may be thoroughly
cleaned in the dishwasher once a week.
Hand Wash Instructions
Step 1
Hand wash the mask in warm water with a mild dishwashing
detergent. Do not use bleach, alcohol, cleaning solutions
containing alcohol, or any strong household cleaners. Do not
use cleaners containing conditioners or moisturizers.
Step 2
Rinse thoroughly. Air dry. Make sure the mask is dry before
use. Inspect the mask after cleaning. Replace the mask if any
parts are damaged or deteriorated.
Dishwasher Instructions
With normal use and proper care, the mask has an expected useful life of six (6) months. Depending on frequency
of use, care, and cleaning, useful life expectancy may be up to one year.
Step 1
Wash the mask in the top shelf of the dishwasher. NOTE: Place the cushion with the rigid side
down to avoid deformation of the cushion. Do not wash the headgear in the dishwasher.
Step 2
Make sure the mask is dry before use. Air dry if necessary. Inspect the mask after cleaning.
Replace the mask if any parts are damaged or deteriorated.
C
aUtIon
! Use liquid dishwashing detergent only.
Headgear with Swivel Clip Mask Accessories
Step 1
Hand or machine wash the headgear and attached swivel clips with a standard laundry deter-
gent and warm water. Rinse thoroughly. Do not use bleach.
Step 2
Line dry or machine dry at the medium heat setting.
NOTE: Slight readjustment of the straps may be needed after repeated
washings.
r
emovIng
tHe
H
eadgear
and
m
ask
You can disconnect the headgear from the mask in the following ways:
• Disconnect the forehead support by pressing on the center of the forehead
support and then gently pressing the sides of the support arm together.
• Disconnect the neck straps by removing the swivel clips from their sockets on
the mask.
• Slide the headgear over your head. The entire headgear stretches.
To quickly release the flexible tubing, press on both tubing quick-release tabs
located on either side of the entrainment valve.
Press center of
forehead support
Press sides of support
arm together
a
djUstIng
tHe
C
omfort
f
Ull
2 m
ask
and
p
remIUm
H
eadgear
1. Check the Entrainment Valve. With the airflow turned
off, verify that the entrainment valve flapper is lying flat
so that room air can flow through the large opening in
the valve. Turn the airflow on; the flapper should now
cover the large opening, and air from the CPAP or bi-level
device should flow into the mask. If the flapper does not
close or does not function properly, replace the mask.
2.
Adjust the StabilitySelector™ of the Support Arm to its
uppermost position. Hold the mask cushion gently over your nose. Slide the
headgear over your head. You may want to disconnect one of the swivel clips
before putting on the headgear. Reconnect the swivel clip after you have the
headgear in position.
3. Connect the flexible tubing (included with your CPAP or bi-level device) to the swivel on the mask.
4. Turn on the airflow.
5. Lie down on your back and breathe normally. You may need to readjust the forehead arm to minimize leaks.
6. Loosen the EZPeel™ tabs to unhook the straps and gradually tighten them until the mask has minimal
leaks and fits comfortably. Do not overtighten the headgear! Overtightening can irritate your face, cause or
worsen leaks, or damage the mask.
W
arnIng
!
A minimum pressure of 3.0 cm H
2
O at the mask (CPAP or EPAP) is required.
W
arnIng
!
Do not block or try to seal the large opening or the exhalation ports on the valve.
Ensure that the valve is not blocked with secretions and that the flapper is dry.
r
eplaCIng
tHe
f
oreHead
p
ad
If you need to replace the forehead pad, contact your home care provider to reorder. Follow the
instructions below.
Step 1 Remove the pad. Slide the pad outward toward
one edge of the forehead support, and lift it
upward so the connection track on that side is
released from the plastic connection slot. Next,
pull the pad in the opposite direction to release
the connection track on the other side from its
connection slot.
Step 2 Install the new pad. Slide one of the connection
tracks on the new pad into its connection slot
on the support. Pull the pad in the opposite direction (toward the center) and slide the other
connection track into its connection slot to lock the pad into place.
Large Opening
Flapper (inside)
Exhalation Ports
Swivel
Connection
Track
Connection
Slot
Forehead
Support
Premium Headgear with EZPeel™ tabs
Forehead Support
Entrainment Valve
(exhalation port)
DO NOT BLOCK
Flexible Tubing
(to therapy device)
Tubing Quick Release
Cushion