44
User Manual
Chapter 10: Accessories
There are several accessories you can use with the device.
10.1 Adding a Humidifier
The REMstar Heated Humidifier, REMstar Pass-over Humidifier, and
H2 Heated Humidifier are available from your home care provider. The
humidifiers may reduce nasal dryness and irritation by adding moisture
(and heat, if applicable) to the airflow.
CAUTION:
For safe operation, the humidifier must always be
positioned below the circuit connection at the mask and
the air outlet on the device. The humidifier must be level
for proper operation.
Refer to the humidifier instructions for complete setup information.
10.2 Adding Oxygen to the Device
Oxygen may be added to the mask connection. Please note the warnings
listed below when using oxygen with the device.
WARNING:
If you are using oxygen, your BiPAP S/T must be
equipped with the Respironics Pressure Valve (Part
number 302418). Failure to use the Pressure Valve could
result in a fire hazard.
WARNING:
Oxygen accelerates fires. Keep the device and the
O
2
containers away from heat, open flames, any oily
substance, or other sources of ignition. Do not smoke in
the area near the device or the O
2
container.
WARNING:
When using oxygen with your device, the oxygen supply
must comply with the local regulations for medical
oxygen.
WARNING:
When using oxygen with this system, turn the device
on
before turning the oxygen on. Turn the oxygen
off
before turning the device off. This will prevent oxygen
accumulation in the device.