Chapter 4: Controls and
Display Features
Figure 4–1 shows the location of the device’s alarm power indicators,
control panel,
Pressure On/Off
button, and breathing circuit connection.
Breathing Circuit
Connection
Control Panel
Pressure On/Off
Button
Alarm and Power Indicators
Figure 4–1 Front and Top
4.1 Pressure On/Off Button
The
Pressure On/Off
button, located on the side of the unit, starts and
stops the unit’s airflow. Press the button in to turn the airflow on. This
puts the device in the
Operate
state. Depress the button to turn the
airflow off and put the device in the
Standby
state.
When the device is in Standby, any ramp in progress is terminated, the
alarms are reset (except for the System Errors alarm), and the humidifier
is turned off.
The
Pressure On/Off
button is independent of the display screen.