29
User Manual
To change the Flex setting, complete the following steps:
1. From either the Monitoring or Standby screens, press the
Right User
button. The Flex Setting screen appears, as shown in Figure 6–8.
Figure 6–8 Flex Setting Screen
2. To increase or decrease the Flex setting, press the
Heat
or
Ramp
button until the correct setting appears. You can choose from
1
to
3
.
NOTE:
It is recommended that you start with the minimum setting of
1, which provides the least relief. Levels 2 and 3 progressively
increase the pressure relief.
6.2.2.2 Changing the Rise Time Setting
Rise time is the time it takes for the device to change from EPAP to IPAP.
You can adjust the rise time to find the setting that provides you with the
most comfort.
NOTE:
The rise time feature is not prescribed for all users. If the
screen shown in Figure 6–9 does not display, you cannot
adjust this setting.
Additionally, if the Flex feature has been prescribed for you,
when Flex is enabled, the rise time is fixed at a setting of 3.
The Rise Time screen will not display, and you won’t be able
to adjust the setting.