36
User Manual
7.3.2 Medium Priority Alarms
Alarm
LED
Alarm
Sound
Display
Message
Device
Action
Possible
Cause
Your
Action
Yellow Flash
• • •
DC Power
LED Flashes
Operates
Battery nearly
discharged.
Press the RESET button to
reset the alarm. Replace
the battery. If the alarm
continues, contact your
home care provider.
7.3.3 Low Priority Alarms
Alarm
LED
Alarm
Sound
Display
Message
Device
Action
Possible
Cause
Your
Action
Yellow Solid
• •
CARD
flashes
and
card error
code ("Cxx")
displays
Operates
There is a problem
with the SmartCard
inserted in the
SmartCard
connectivity slot.
Perhaps the
SmartCard is
inserted upside down
or backwards.
Confirm that the
SmartCard is properly
inserted.
If the alarm continues
to occur, remove the
SmartCard from the
device and contact
your home care
provider.
Operates
Device lost AC
power and is now
operating on DC
power.
Press the RESET
button to reset the
alarm. Check the
AC power. Seek a
reliable power
source. Provide AC
power if you do not
want to use a battery;
otherwise, no further
action is needed.
At start-up only,
alarm notifies you
that a battery is
being used
to provide power.
Yellow Solid
DC power
LED flashes
• •
Unchanged
Operates
Operates
Yellow Solid
AC power
LED flashes
Yellow Solid
• •
• •
The device
has successfully
downloaded the
prescription from
the SmartCard.
Remove the
SmartCard from the
device. If alarm
continues to occur,
contact your home
care provider.
ALARM,
CARD and
cm H
2
O
symbols
flash
The DC output
voltage from the AC
power supply is out
of spec (< 22 VDC)
or there is a
defective battery
sense line on the
DC power adapter.
Remove AC power
from the power supply
and then restore
power. If alarm
continues to occur,
contact your home
care provider.