11.3 Drawings and Schematics
11.3.1Battery Pack Electrical Schematic
Votage Readings
Using a standard Digital Voltmeter, the following voltage
readings will be obtained from a fully charged battery:
Pin 1 to Pin 4 = 7.0 VDC (± depending on polarity of leads)
Pin 5 to Pin 7 = 7.0 VDC (± depending on polarity of leads)
Pin 5 to Pin 6 = 7.0 VDC (± depending on polarity of leads)
Pin 6 to Pin 7 = 0.0 VDC
NOTE: The battery packs have three (3) contacts on one end
and four (4) contacts on the other end. Holding the battery pack
with the contacts towards you, the contacts are numbered from
left-to-right.