BiliChek Service Manual
19
1014988
A
Hand Held Unit (HHU)
Press and release either
button on top of the HHU
(
F1
or
F2
).
Does the
display come
on?
No
T r y a n o t h e r
Battery pack.
B
Yes
R ec or d t h e di sp lay
information.
R
e
f
e
r
t
o
T r o u b l e s h o o t i n g
guide, Section 11.1
Will the HHU
calibrate
T a k e 5 s k i n
measurements.
Yes
OK ?
Yes
The HHU is operating
correctly, and can be
returned to use.
No
T r y s e v e r a l t i m e s . I f s t i l l n o
calibration , remove and rotate the
Bili
Cal
tip and try again to calibrate. If
the error is still present, change the
meas urement ligh t bulb and try to
calibrate.
NOTE:
It will take several
t ry s t o c ali brate af te r a l igh t bu lb
change. If the error is still present,
r e c or d t h e e r r o r / s u b e r r o r c o de
information found in section 11.0 and
contact your dis tributor to reques t
i n s t r u c t i o n s f o r r e t u r n i n g t h e
defective HHU to Respironics for repair.
If the H HU will not tak e
s k i n m e a s u r e m e n t s ,
con tac t your distribut or
and request instruc tions
for returning the defective
H H U t o Respironics f o r
repair.
No
B
P l a c e a b a t t e r y
p a c k i n t h e
c h a r g er b as e i n
the right hand well.
Well "B".
Does the right hand
BSI light glow ?
No
P l a c e a b a t t e r y
pac k in the HH U
and place the HHU
i n t h e c h a r g e r
b a s e l e f t h a n d
well. Well "A"
Does the left hand
BSI light glow ?
Yes
O bs er ve t h e co lor of
indicator:
Amber = charging
Green = Fully charged
Red = Defective Battery
pack
(See page 12 for amplification)
Charger Base
Remove all battery packs
and unplug the charger
bas e. R ec on nect th e
charger base to power
while observing the BSI
lights.
T h e B S I l i g h t
s eq u e n c e s h o u l d
be:
BOTH
LIGHTS
- R e d - G r e e n -
Amber - Off
OK ?
No
T he c har ger
b a s e o r t h e
power supply
is defective.
C
C
M ake su re there is a
fully
charged
battery pack in the
HHU
Yes
Does
the MSI light glow ? See
section 10.0 of this
manual
Yes
N O T E :
A f t er t h e s ec on d
battery pack try, check the
battery pack output with a
mu l t i met er . I t s h o u ld be
between 7 and 7.5 Volts DC.
N O T E :
I f t h e di s pl ay is
f r a g m e n t e d ( s e g m e n t s
m i s s i n g ) , t h e H H U i s
def ec t iv e. C on t ac t y o u r
distributor for instructions on
the return of defective HHU
to Respironics for repair..
NOTE
: The indicator light will
t ak e abou t 3 s ec o n ds t o
illuminate.
No
D
I n s e r t a b a t t e r y
p a c k i n t o t h e
charger base right
hand well. Well "B".
Yes
11.2 Troubleshooting Flow Chart
Summary of Contents for bilichek
Page 1: ...Service Manual www respironics com ...
Page 23: ...BiliChek Service Manual 23 1014988 11 3 3 Charger Base PCB Schematic ...
Page 27: ...BiliChek Service Manual 27 1014988 11 3 4 Charger Base PCB Component Layout ...
Page 36: ...Reorder B800 25 Revision 1014988 DS 07 14 03 www respironics com ...