BiliChek Service Manual
28
1014988
11.4 Hand Held Unit (HHU) Error Code List with Explainations
NOTE:
All sub errors are displayed in the date field unless specified otherwise in each section. See Illustration at the top of
page 32.
Error: E01 Calibration
Sub Errors Description
000001
Reference component out of range. Install a fully charged battery pack and try several calibrations with
new Bili
Cal
tips. If error persists, HHU requires service.
000002
Dynamic integration time out of range. Install a fully charged battery pack and try several calibrations
with new Bili
Cal
tips. If error persists install a new measurement light bulb and try calibration several
times. If the error still persists, the HHU requires service.
000004
Shape index above threshold. Try a fully charged battery pack and a new Bili
Cal
tip. If error persists,
HHU requires service.
000008
Peak pixel not found. Install a fully charged battery pack and try several calibrations with new Bili
Cal
tips. If error persists, HHU requires service.
008192
Stray Light Slope Error. Install a fully charged battery pack and try several calibrations with new Bili
Cal
tips. If error persists, HHU requires service.
016384
Mean of dark scan data above threshold. Install a fully charged battery pack and try several calibrations
with new Bili
Cal
tips. If error persists, HHU requires service.
NOTE: Firware
Version
≥
2.4/3.8
Only
Summary of Contents for bilichek
Page 1: ...Service Manual www respironics com ...
Page 23: ...BiliChek Service Manual 23 1014988 11 3 3 Charger Base PCB Schematic ...
Page 27: ...BiliChek Service Manual 27 1014988 11 3 4 Charger Base PCB Component Layout ...
Page 36: ...Reorder B800 25 Revision 1014988 DS 07 14 03 www respironics com ...