BiliChek Service Manual
32
1014988
12.0 Routine (Operator) Cleaning
The Bili
Chek
System is made of an easy-to-clean material
with a smooth, textured surface. One of the following clean-
ing agents is recommended for cleaning the exterior of the
HHU, Battery Charger Base, and Battery Packs:
♦
KleenaSeptic
®
♦
Cavicide
®
♦
70 to 90% Isopropyl Alcohol solution
♦
1% Bleach solution
Spray the cleaning agent on a soft damp cloth and wipe clean
the Bili
Chek
system. The same procedure can be used to
clean the display window.
Caution:
Clean the measurement tip with 90% or higher
Isopropyl Alcohol using a soft optical surface cleaning wipe.
WARNING:
Do not immerse any part of the system in water
or other liquid. If liquids spill onto the unit, wipe with a damp
cloth and allow to dry prior to using the system.
WARNING:
Do not attempt to clean and/or reuse the
Bili
Cal
Individual Calibration Tip.
13.0 Maintenance and Repair and
Replacement
13.1 Hand Held Unit (HHU) Measurement Light Bulb
Replacement
If the instrument bulb is not working and error code E01 or
E10 (refer to the troubleshooting section 11.4) is displayed, it
is necessary to replace the measurement light bulb. A Re-
placement Bulb Kit (RI P/N B800-22) is available from
Respironics or your local distributor.
WARNING:
The Bili
Chek
®
Non-Invasive Bilirubin
Analyzer is to be used only with the approved Replacement
Bulb Kit from Respironics. Do not over-tighten the lamp
cap.
♦
Remove the battery pack from the bottom of the HHU.
♦
Remove the protective label covering the lamp
assembly.
♦
Remove the lamp cap using a flat tip screwdriver.
♦
Grab the loop of the light bulb with forceps or similar
tool/instrument and gently pull to remove the lamp
assembly.
♦
Insert the replacement lamp assembly using caution
not to touch bulb with fingers. Be sure to place bulb
into socket firmly. DO NOT use the lamp
cap to tighten bulb into place.
♦
Apply the new lamp cap and protective label furnished
in the kit.
♦
Insert a fully charged battery pack into the bottom of
the HHU.
Summary of Contents for bilichek
Page 1: ...Service Manual www respironics com ...
Page 23: ...BiliChek Service Manual 23 1014988 11 3 3 Charger Base PCB Schematic ...
Page 27: ...BiliChek Service Manual 27 1014988 11 3 4 Charger Base PCB Component Layout ...
Page 36: ...Reorder B800 25 Revision 1014988 DS 07 14 03 www respironics com ...