1054417, V
ER
. 01
Alice
®
PDx
TM
Service & Technical Reference Manual
4. Slightly separate the Display Enclosure from the Display Enclosure Cover (part of the Clamshell
Assembly).
5. Unlock the J10 connector on the Digital PCA (part of the Main PCA Stack) and remove the ribbon
cable. The ribbon cable is connected to the patient event button installed in the Display Enclosure.
F
IGURE
G: D
ISPLAY
E
NCLOSURE
R
EMOVAL
/I
NSTALLATION
CAUTION
A Ribbon Cable is connected between the Patient Event Button in the Display Enclosure and
the Main PCA Stack. Slightly separate the Display Enclosure from the Clamshell Assembly,
then disconnect the Ribbon Cable from the J10 connector on the Digital PCA.
Unlock this connector
(J10) to disconnect
the ribbon cable.
Ribbon cable
Display Enclosure
IMPORTANT - DURING ASSEMBLY,
THIS RIBBON CABLE MUST BE
ROUTED AND FOLDED AS SHOWN.
BE SURE THAT IT IS NOT
PROTRUDING FROM THE MAIN PCA
STACK.