F
IGURE
L: D
ISPLAY
H
OLDER
NOTE
The Hinge Cup shown in Figure L replaces a piece of foam
that was installed in earlier versions of the Alice PDx.
Remove foam from LCD. Do not
touch LCD with bare fingers.
Snap Display into Display
Enclosure Cover
Be sure the Flex Cable does not get
caught or pinched between plastic parts.
Remove protective Film
inside of Display Cover
Position Hinge Cup in
the Enclosure Cover
Connect Sensor Tube
to Cannula Port