Alice
®
PDx
TM
Service & Technical Reference Manual
R
EPLACING
THE
C
ANNULA
P
ORT
To remove the Cannula Port:
1. Remove the Display Enclosure. Refer to page 36 if necessary.
2. Remove the Main PCA Stack. Refer to page 39 for removal instructions if necessary.
3. Unsnap the Cannula Port Cover from the latches on the Cannula Port mount.
4. Remove the Cannula Port.
F
IGURE
M: C
ANNULA
P
ORT
R
EMOVAL
/I
NSTALL
To install the Cannula Port:
1. Place the Cannula Port the opening in the Display Enclosure as shown in Figure M.
2. Place the Cannula Port cover over the Cannula Port and snap it into place to secure the Cannula
Port.
3. Assemble the remainder of the Alice PDx as necessary.
Cannula Port Cover
Cannula Port
Place Cannula Port through opening in
Display Enclosure Cover, then snap the
Cannula Port Cover onto the Cannula Port.
Display Enclosure Cover