Problem
Possible cause
Solution
Flow generator does not
start when you press the
Start/Stop
or
Delay
Timer
button (and the
LCD screen is blank)
Power cord is not connected properly
Check both ends of the power cord
Flow generator is not switched on
Switch on main switch at rear of unit
Power outlet may be faulty
Try another power outlet
SmartStart is enabled but
flow does not start when
you begin breathing in
the mask
Breath is not deep enough to trigger
SmartStart
Take a deep breath in and out
through the mask
Mask, mask/valve or mask/humidifier
combination interfering with
SmartStart triggering
Use
Start/Stop
button
Excessive leak
•Adjust position of mask and
headgear
•Plugs may be missing from ports on
mask; replace them
•Air tube not connected properly;
connect air tube firmly at both ends
•Air tube kinked or punctured;
straighten or replace air tube
SmartStart not enabled (
does not
appear on the LCD screen)
Use
Start/Stop
button
Air leaking around nose
and eyes
Nasal cushion not fitted properly onto
mask frame
Fit cushion properly
Mask and headgear straps not
positioned correctly
Adjust position of mask and headgear
(Note: tighter is not always better)
Plugs missing from mask ports
Replace plugs
Nasal cushion torn or distorted
Replace cushion
SmartStart is enabled,
but the flow generator
does not stop
automatically when you
remove your mask
SmartStart not enabled (
does not
appear on the LCD screen)
Use
Start/Stop
button
Mask, mask/valve or mask/humidifier
combination interfering with
SmartStart triggering
Use
Start/Stop
button
Mask alarm enabled (
M
appears on
the LCD screen)
(continued over page)
SmartStart will not stop the flow
automatically when the Mask Alarm is
enabled; use the
Start/Stop
button
Summary of Contents for VPAP II
Page 1: ...VPAP II II ST Reorder number 218038 2 00 03...
Page 4: ...1 E NGLISH VPAP II II ST USER S MANUAL RESMED...
Page 15: ...12...
Page 17: ...14...
Page 25: ...22...
Page 28: ...25 D EUTSCH VPAP II II ST BEDIENUNGSANLEITUNG RESMED...
Page 33: ...30...
Page 43: ...40...
Page 51: ...48...
Page 56: ...53 F RAN AIS VPAP II II ST MANUEL DE L UTILISATEUR RESMED...
Page 61: ...58...
Page 71: ...68...
Page 79: ...76...
Page 81: ...78...
Page 84: ...81 I TALIANO VPAP II II ST MANUALE DELL UTENTE RESMED...
Page 89: ...86...
Page 99: ...96...
Page 107: ...104...
Page 109: ...106...
Page 112: ...109 E SPA OL VPAP II II ST MANUAL DEL USUARIO RESMED...
Page 117: ...114...
Page 127: ...124...
Page 135: ...132...
Page 140: ...137 P ORTUGU S VPAP II E II ST MANUAL DO USU RIO RESMED...
Page 145: ...142...
Page 155: ...152...
Page 163: ...160...
Page 165: ...162...
Page 168: ...165 S VENSKA VPAP II II ST BRUKSANVISNING RESMED...
Page 181: ...178...
Page 189: ...186...
Page 191: ...188...