background image

 

English 5 

 

 CAUTION 

Do not use the device with water in the humidifier on a plane due to the risk of inhalation 
of water during turbulence. 

 

Troubleshooting 

If you have any problems, have a look at the following troubleshooting topics. If you are not 
able to fix the problem, contact your care provider or ResMed. Do not try to open the device. 

 

General troubleshooting 

Problem/possible cause 

Solution 

I am getting a dry or blocked nose 

Humidity level may be set too low. 

Adjust the Humidity Level. 

If you have ClimateLineAir heated air tubing, see the 
ClimateLineAir user guide. 

I am getting droplets of water on my nose, in the mask and air tubing 

Humidity level may be set too high. 

Adjust the Humidity Level. 

If you have ClimateLineAir heated air tubing, see the 
ClimateLineAir user guide. 

My mouth is very dry and uncomfortable 

Air may be escaping through your mouth. 

Increase the Humidity Level. 

You may need a chin strap to keep your mouth closed or a 
full face mask. 

 

 
 

Summary of Contents for HumidAir

Page 1: ...User guide English...

Page 2: ...es heated humidification and is designed for use with your AirSense 10 or AirCurve 10 device Read this guide together with your AirSense 10 or AirCurve 10 device user guide Setup CAUTION Do not overfi...

Page 3: ...turn on therapy or change settings refer to your device user guide Recommended water type France only The following water types are recommended for use in the humidifier Boiled tap water cooled to ro...

Page 4: ...outh or moisture in your mask consider using ClimateLineAir heated air tubing ClimateLineAir together with Climate Control delivers more comfortable therapy Removing the humidifier 1 Hold the humidifi...

Page 5: ...regularly check the humidifier for any damage Replace it if it is leaking or has become cracked cloudy or pitted Replace it if the seal is cracked or torn Remove any white powder deposits using a solu...

Page 6: ...cause Solution I am getting a dry or blocked nose Humidity level may be set too low Adjust the Humidity Level If you have ClimateLineAir heated air tubing see the ClimateLineAir user guide I am getti...

Page 7: ...er guide for fitting instructions or use the Mask Fit function to check your mask fit and seal System fault refer to user guide Error 004 Device may have been left in a hot environment Allow to cool b...

Page 8: ...ly come off in order to avoid damage to the parts or the device You can easily reassemble them as described below To insert the humidifier seal 1 Place the seal into the lid 2 Press down along all edg...

Page 9: ...ts side as water might get into the device Do not use bleach chlorine alcohol or aromatic based solutions moisturising or antibacterial soaps or scented oils to clean the device the humidifier or air...

Page 10: ...Sea level to 2 591 m air pressure range 1013 hPa to 738 hPa Storage and transport temperature 20 C to 60 C Storage and transport humidity 5 to 95 relative humidity non condensing Physical device and...

Page 11: ...e following symbols may appear on the product or packaging Indicates a warning or caution Read instructions before use Manufacturer European Authorised Representative Batch code Catalogue number Seria...

Page 12: ...om for other ResMed locations worldwide For patent information see www resmed com ip AirSense AirCurve HumidAir ClimateLine and SlimLine are trademarks of ResMed Ltd AutoSet SlimLine and ClimateLine a...