Humidifier performance 

The following settings have been tested at 71.6°F (22°C) ambient temperature: 

Mask Pressure 
cm H

2

O (hPa) 

RH output % 

Nominal system output AH

1

, BTPS

2

 

Setting 4 

Setting 8 

Setting 4 

Setting 8 

3 85 100 

6 >10 

4 85 100 

6 >10 

10 85 100 6  >10 
20 85 90 6  >10 
25 85 80 6  >10 

1

 AH - Absolute Humidity in mg/L 

2

 BTPS - Body Temperature Pressure Saturated 

 

Symbols 

The following symbols may appear on the product or packaging. 

 

 Indicates a warning or caution. 

 Read instructions before use. 

 Manufacturer. 

 European Authorised Representative. 

 Batch code. 

 Catalogue number. 

 Serial number. 

 Protected against finger sized objects and against dripping water 

when tilted up to 15 degrees from specified orientation. 

 Humidity limitation. 

 Temperature limitation. 

 Maximum water level. 

 Use distilled water 

only. 

 European RoHS. 

 China pollution control logo 1. 

 China pollution control 

logo 2. 

 Operating altitude. 

 

Limited warranty 

Refer to your device user guide. 

 

Summary of Contents for HumidAir

Page 1: ...User guide English...

Page 2: ...es heated humidification and is designed for use with your AirSense 10 or AirCurve 10 device Read this guide together with your AirSense 10 or AirCurve 10 device user guide Setup CAUTION Do not overfi...

Page 3: ...turn on therapy or change settings refer to your device user guide Recommended water type France only The following water types are recommended for use in the humidifier Boiled tap water cooled to ro...

Page 4: ...outh or moisture in your mask consider using ClimateLineAir heated air tubing ClimateLineAir together with Climate Control delivers more comfortable therapy Removing the humidifier 1 Hold the humidifi...

Page 5: ...regularly check the humidifier for any damage Replace it if it is leaking or has become cracked cloudy or pitted Replace it if the seal is cracked or torn Remove any white powder deposits using a solu...

Page 6: ...cause Solution I am getting a dry or blocked nose Humidity level may be set too low Adjust the Humidity Level If you have ClimateLineAir heated air tubing see the ClimateLineAir user guide I am getti...

Page 7: ...er guide for fitting instructions or use the Mask Fit function to check your mask fit and seal System fault refer to user guide Error 004 Device may have been left in a hot environment Allow to cool b...

Page 8: ...ly come off in order to avoid damage to the parts or the device You can easily reassemble them as described below To insert the humidifier seal 1 Place the seal into the lid 2 Press down along all edg...

Page 9: ...ts side as water might get into the device Do not use bleach chlorine alcohol or aromatic based solutions moisturising or antibacterial soaps or scented oils to clean the device the humidifier or air...

Page 10: ...Sea level to 2 591 m air pressure range 1013 hPa to 738 hPa Storage and transport temperature 20 C to 60 C Storage and transport humidity 5 to 95 relative humidity non condensing Physical device and...

Page 11: ...e following symbols may appear on the product or packaging Indicates a warning or caution Read instructions before use Manufacturer European Authorised Representative Batch code Catalogue number Seria...

Page 12: ...om for other ResMed locations worldwide For patent information see www resmed com ip AirSense AirCurve HumidAir ClimateLine and SlimLine are trademarks of ResMed Ltd AutoSet SlimLine and ClimateLine a...

Reviews: