2

 

 

 

1.

 

Place the device on a stable level surface. 

2.

 

Remove the side cover from the device by pressing the clip of the side cover and pulling it 
away from the device. 

3.

 

Open the humidifer and fill it with water up to the maximum water level mark. 
Do not fill the humidifier with hot water. 

4.

 

Close the humidifier and insert it into the side of the device. 
Your device is now ready to be used with humidification. See 

Humidity Level

 to change 

your settings. 

 

For more information on how to setup your device, turn on therapy or change settings refer to 
your device user guide. 

 

Recommended water type

 (France only) 

The following water types are recommended for use in the humidifier: 

 

Boiled tap water, cooled to room temperature 

 

Distilled water 

 

Bottled water (with a low mineral content). 

Note

: Do not use non-sterile demineralised water for domestic use. 

 

Humidity Level 

The humidifier moistens the air and is designed to make therapy more comfortable. If you are 
getting a dry nose or mouth, turn up the humidity. If you are getting any moisture in your mask, 
turn down the humidity. 

 

You can set the Humidity Level to Off or between 1 and 8, where 1 is the lowest humidity 
setting and 8 is the highest humidity setting. 

 

Summary of Contents for HumidAir

Page 1: ...User guide English...

Page 2: ...es heated humidification and is designed for use with your AirSense 10 or AirCurve 10 device Read this guide together with your AirSense 10 or AirCurve 10 device user guide Setup CAUTION Do not overfi...

Page 3: ...turn on therapy or change settings refer to your device user guide Recommended water type France only The following water types are recommended for use in the humidifier Boiled tap water cooled to ro...

Page 4: ...outh or moisture in your mask consider using ClimateLineAir heated air tubing ClimateLineAir together with Climate Control delivers more comfortable therapy Removing the humidifier 1 Hold the humidifi...

Page 5: ...regularly check the humidifier for any damage Replace it if it is leaking or has become cracked cloudy or pitted Replace it if the seal is cracked or torn Remove any white powder deposits using a solu...

Page 6: ...cause Solution I am getting a dry or blocked nose Humidity level may be set too low Adjust the Humidity Level If you have ClimateLineAir heated air tubing see the ClimateLineAir user guide I am getti...

Page 7: ...er guide for fitting instructions or use the Mask Fit function to check your mask fit and seal System fault refer to user guide Error 004 Device may have been left in a hot environment Allow to cool b...

Page 8: ...ly come off in order to avoid damage to the parts or the device You can easily reassemble them as described below To insert the humidifier seal 1 Place the seal into the lid 2 Press down along all edg...

Page 9: ...ts side as water might get into the device Do not use bleach chlorine alcohol or aromatic based solutions moisturising or antibacterial soaps or scented oils to clean the device the humidifier or air...

Page 10: ...Sea level to 2 591 m air pressure range 1013 hPa to 738 hPa Storage and transport temperature 20 C to 60 C Storage and transport humidity 5 to 95 relative humidity non condensing Physical device and...

Page 11: ...e following symbols may appear on the product or packaging Indicates a warning or caution Read instructions before use Manufacturer European Authorised Representative Batch code Catalogue number Seria...

Page 12: ...om for other ResMed locations worldwide For patent information see www resmed com ip AirSense AirCurve HumidAir ClimateLine and SlimLine are trademarks of ResMed Ltd AutoSet SlimLine and ClimateLine a...

Reviews: