RMP60 installation guide
7.4
Fault-finding
Symptom
Cause
Action
RMP60 status LEDs do
not correspond to RMI-Q
or RMI-QE
status LEDs.
Radio link failure – RMP60 out of
RMI-Q or RMI-QE range.
Check position of RMI-Q or
RMI-QE, see operating envelope.
RMP60 has been enclosed/
shielded by metal.
Review installation.
RMP60 and RMI-Q or RMI-QE
are not partnered.
Partner RMP60 and RMI-Q or
RMI-QE.
RMI-Q or RMI-QE error
LED lit during probing
cycle.
Probe not switched on or probe
timed out.
Change setting. Review turn off
method.
Probe out of range.
Check position of RMI-Q or
RMI-QE, see operating envelope.
Dead batteries.
Change batteries.
RMP60 and RMI-Q or RMI-QE
are not partnered.
Partner RMP60 with RMI-Q or
RMI-QE.
Probe selection error.
Verify that one radio probe is
working and is correctly selected
on the RMI-Q or RMI-QE.
0.5 second turn-on error.
Ensure that all radio probes are
‘Q’ or ‘QE’ marked probes, or
change the RMI-Q or RMI-QE
turn-on time to 1 second.
RMI-Q or RMI-QE low
battery LED lit.
Low batteries.
Change batteries soon.
Reduced range.
Local radio interference.
Identify and remove.
Summary of Contents for RMP60
Page 1: ...RMP60 QE radio machine probe Installation guide H 6587 8520 01 A...
Page 6: ...RMP60 installation guide iv Contents This page is intentionally left blank...
Page 12: ...RMP60 installation guide 1 6 Before you begin This page is intentionally left blank...
Page 24: ...RMP60 installation guide 2 12 RMP60 basics This page is intentionally left blank...
Page 64: ...RMP60 installation guide 7 6 Fault finding This page is intentionally left blank...
Page 67: ......