Fixing FA10 calibration artefact to CMM
The FA10 calibration artefact is designed to be fixed securely to the bed of the CMM to allow accurate and repeatable calibration of RFP1. The
diagrams below show how to fix the FA10 to the bed of the CMM.
1. Locate an appropriate threaded hole in the bed of the machine.
2. Position the artefact plate above the threaded hole.
3. Screw the correct cap head bolt through the hole into the threaded hole on the machine. The supplied washer can be used to correctly fit
smaller diameter bolts.
4. Tighten with a hex wrench.
5. Attach the calibration pillar to the kinematic mount. Integral magnets will secure it in place. Polarisation of the magnets will ensure the
pillar is fitted in the correct orientation.
RFP1 user's guide
www.renishaw.com
Issued 06 2022
20