Installation
NOTE:
Please refer to the REVO-2 change port system spacing guide (Renishaw part number H-1000-5408) for full details of the
recommended spacing for positioning all sensors and artefacts on the MRS / MRS2 rail.
Fitting VPCP to the MRS / MRS2 rack
The VPCP heated change ports can be fixed to an MRS or MRS2 rack system. It is recommended that they are attached to the MRS / MRS2 rail
using the following procedure, where it is assumed that the MRS / MRS2 rack system is correctly installed.
1. Insert one of the fixing screws through the VPCP.
2. Position the VPCP underneath the rail and locate the respective T-nut within the rail*.
3. Hand tighten the fixing screw into the T-nut and repeat the process for the next fixing screw.
4. Position the VPCP and tighten both fixing screws using the hexagonal key supplied.
* NOTE:
T-nuts must be used with the MRS system. However T-nuts and D-nuts are compatible with the MRS2 system.
RFP1 user's guide