![Renishaw PI 200 Installation Manual Download Page 7](http://html1.mh-extra.com/html/renishaw/pi-200/pi-200_installation-manual_1441482007.webp)
Safety
Safety
If this product is not used in its intended manner, any protection provided may be
impaired.
There are no user serviceable parts inside this equipment.
The PI 200 interface unit must be connected to a supply incorporating a protective
earth conductor via a three-core mains cable (line cord).
Electrical ratings
Supply voltage range
100 - 240 V ac + 10%, -15%
Power frequency range
47 Hz - 63 Hz
Power consumption
10 W
SSR contact ratings
±50 V pk.
±40 mA pk.
Operating conditions
The PI 200 interface unit is specified to operate under the following conditions as
defined in BS EN 61010-1:2001.
Protection provided by
enclosure
IP30
Altitude
Maximum 2000 m
Operating temperature
0 °C to +50 °C
Storage temperature
-10 °C to +70 °C
Relative humidity
Maximum 80% RH up to +31 °C, decreasing
linearly to a maximum 50% at +40 °C.
Transient overvoltage
Installation category II
Pollution degree
2
The PI 200 is isolated from AC power by disconnection of the IEC mains connector on
the rear panel. If any additional means of isolation is required, it must be specified and
fitted by the machine manufacturer or the installer of the product. The isolator must
be sited within easy reach of the CMM operator and comply with IEC61010 and any
applicable national wiring regulations for the country of installation.