2
Installation procedure
Installation procedure
WARNING:
Ensure the PI 200 is disconnected from the mains supply during
installation.
WARNING:
Take care not to exceed the operation ambient of 50 °C around
the unit. Do not install near sources of heat. Forced cooling may be required
in final installation.
. Rack mounting
The PI 200 may be mounted in a standard 19 in equipment rack either alone or
alongside a Renishaw controller as indicated in figures 13 and 14.
POWER
TYPE
PROBE
RESET
ON
STD
TP200 SEATED DAMPED STOP
PROBE INTERFACE
R
Figure -
Mounting the PI 200
To mount the PI 200 as indicated in figure 13 requires 2 x 1/3 blanking panel kits (part
number A-1018-0123). Remove the blanking plugs from both sides of the PI 200 and fit
the panels using the M5 x 6 mm long screws provided.
A-1018-0123
Figure -
PI 200 with PHC10-2
To mount the PI 200 adjacent to a Renishaw controller, as indicated in figures 14 and
15, the following parts are required:
• Rack mounting bracket - part number A-1018-0124
• Enclosure link bracket - part number A-1018-0126
!
!