Product description
Switch :
Inverts the polarity of the HALT output.
Switch 2:
Selects control of the LEDOFF output to either internal PI 200 control according to the
status of SYNC, or sets the output to high impedance for control of the head LED by
external switching.
Switch :
Sets the PI 200’s response to STOP when asserted from an external source. When
enabled, the PI 200 will assert SYNC, HALT and open the SSR relay.
Switch :
Inverts the polarity of the SYNC and SSR relay outputs.
Switch 5:
When enabled, a tone will sound for approximately 160 milliseconds when the probe
triggers and SYNC is asserted. When disabled, the tone will activate only when the
stylus has remained deflected for 10 seconds.
Switches and 7:
These switches select the debounce time and mode of operation as indicated in
figure 8.
Switch :
Overrides the settings of switches 6 and 7, reducing the trigger and reseat debounce
times to less than 2 milliseconds.
Switch :
The probe signal filter has a 300 µs time constant to reduce the effects of stylus
vibration when the stylus makes contact with the workpiece. Renishaw recommends
that this switch is set to the UP position for normal use.