![Renishaw PI 200 Installation Manual Download Page 39](http://html1.mh-extra.com/html/renishaw/pi-200/pi-200_installation-manual_1441482039.webp)
Installation procedure
. Using TP200 with the ACR autochange system
Figure 22 -
Shock absorbing pad fitted to PH10M
The CMM controller must assert the probe damping signal (PDAMP) on the PICS port
when the probe is approaching the rack, to prevent a false trigger being generated
when the probe contacts the docking port lid.
If this is not possible, a shock absorbing pad (part number A-1085-0294 supplied
with ACR1) must be fitted to the autojoint connector on the probe head at the point of
contact with the port lid. Refer to figure 22.
A
0
105