26
Sight glass (4)
clean
(change in case of defect)
After 2 months
Ventilating fan filter
(machine’s back side)
Clean
Cooling water or recirculation
unit
Change water of recirculation
unit, clean the basin
Crucible and muffle support
plates
Check, if necessary change
Water inlet filter (connection
water inlet)
Clean
Compressed air unit
Drain and check the system.
After 6 months
Sinter filter (29) left side of the
device
Unscrew and clean / change
Air filter (30)
Check for contamination
Open filter chamber and
change filter
After 12 months
Thermocouple:
fu mechanic control
change
(Recommendation: Availability
of one thermocouple as spare
part)
2 years
(at least after1500 castings)
Check – up
Complete checkup of the
machine incl. loan system –
Exchange of the worn parts at
a fixed price from REITEL
Note: Water in the air filter is considered to be an indication for humid compressed air. Pls.
have your compressor checked by an expert. Additionally pls. connect a maintenance unit
(REITEL accessories).