36
11625-101-Rev. L
Troubleshooting
Due to the Phoroptor being enclosed and the mechanical nature of the unit, problems associated with
the Phoroptor should be handled by trained Reichert personnel. Please contact Reichert customer
service with the contact information located in the
Introduction
section of this manual, or on the back
cover.
For user serviceable problems, please consult the following list:
Light Does Not Turn On.
(Illuminated Phoroptor
Only.)
Check to see if the male connector is properly inserted into the fe-
male connector on the mounting bracket.
Ensure that the power adaptor is installed into an outlet with the cor-
rect input voltage.
Light may be turned all the way down.
Turn the Rheostat Control knob all the way up.
Weak Sphere Dial Too
Loose or Tight.
Tighten or loosen the coordinating screw on the bottom of the
phoroptor using a standard Allen wrench (not provided.)
Be careful not to adjust the screw too tight or too loose.
Refer to the
Maintenance
section.
Cylinder Power Knob
Too Loose or Tight.
Tighten or loosen the coordinating screw on the bottom of the
phoroptor using a standard Allen wrench (not provided.)
Be careful not to adjust the screw too tight or too loose.
Refer to the
Maintenance
section.
Auxiliary Lens Scale Too
Loose or Tight.
Tighten or loosen the coordinating screw on the bottom of the
phoroptor using a standard Allen wrench (not provided.)
Be careful not to adjust the screw too tight or too loose.
Refer to the
Maintenance
section.