background image

17

11625-101-Rev. L

Optical Corneal Aligning Device

The Phoroptor provides an optically additive lens system and an optical Corneal Aligning Device, both 

essential for a true additive effective power determination.

1.  The additive lens system refers to the addition of lens powers within the Phoroptor. 

2.  The effective power combination of two or more lenses cannot be obtained accurately by simple ad-

dition of their individual powers. 

3.  Allowances must  be made which depend in an intricate way on the powers, the curves, the thick-

nesses, the index of glass, and the air space separating the lenses. 

4.  Accordingly, in the Phoroptor, two essential features have been incorporated to ensure the accuracy 

of the lens prescription.

•  Specially computed lenses and lens separations such that their designated powers can simply 

be added together to give the effective power of any possible combination.

• 

A means for placing this additive lens power system at a specified distance from the eye.

5.  When either of these elements is neglected, the corrective accuracy of the lens system is impaired, 

particularly with regard to combinations of high power lenses.

6.  The distance at which the spectacle lens is generally worn is considered to be 13.75 mm from the 

apex of the cornea to the ocular surface of the lens. 

7.  With this as the standard, the posterior lens surface of the Phoroptor must be placed at a distance of 

13.75 mm if the Phoroptor reading is to be directly applied to spectacle lens power.  

8.  In the Phoroptor, this condition is obtained when the zero setting of the sight in the Corneal Aligning 

Device is lined up with the apex of the cornea.  Refer to Figure OC-1.

9.  To establish proper distance between the patient’s eyes and the instrument, adjust the position of the 

Forehead Rest using the knurled Forehead Rest Knob.

Note: 

Make certain that the patient’s forehead is resting firmly against the headrest. 

 

10. This adjustment will move the patient’s eyes nearer to or farther from the instrument.

11. From the front of the instrument, look into the Corneal Aligning Device. The upper and lower pointers 

should be in exact alignment with the solid black line visible on the mirror.  This is the zero point indi-

cating a 13.75 mm distance from the apex of the patient’s cornea.  Also visible are three hash marks, 

each representing 2 mm additive distance.  Refer to Figure OC-2.

12. With the patient’s forehead positioned against the headrest, adjust the headrest to position the apex 

of the cornea at the zero line (13.75 mm from the lenses). 

13. If, with the headrest retracted, the apex of the cornea appears nasally from the zero line, simply add 

the distance to 13.75 mm. (This figure is the total distance from the cornea to the strong sphere, or 

the vertex distance.)

 

 

Instructions for Use

 (continued)

Figure OC-2   Corneal Alignment 

Device Close Up

Zero 

Point

2mm

4mm

6mm

Figure OC-1   Corneal Alignment Device

Refractionist’s 

Eye

Patient’s 

Eye

Corneal 

Alignment 

Device

Summary of Contents for PHOROPTOR

Page 1: ...REICHERT PHOROPTOR Refracting Instrument User s Guide...

Page 2: ...anges in the product described in this manual without notice and without incorporat ing those changes in any products already sold ISO 9001 13485 Certified Reichert products are designed and manufactu...

Page 3: ...Preparation for Refraction 16 Optical Corneal Aligning Device 17 Taking Measurements 20 Distance Tests 20 Near Tests 21 Operation of the Sphere Lens Dials 22 Cylinder Power and Axis 24 Auxiliary Lens...

Page 4: ...IS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH OR DAMAGE TO THE INSTRUMENT AND OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR WARNING ENSURE THAT THE VOLTAGE APPLIED TO TH...

Page 5: ...IT MAY OCCUR CAUTION DO NOT AUTOCLAVE OR DISINFECT THE UNIT OR ACCESSORIES USING HIGH TEMPERA TURES EXCEEDING THE RECOMMENDED TEMPERATURES INDICATED IN THE SPECIFICATIONS SECTION OF THIS MANUAL OR DAM...

Page 6: ...ropean safety requirements Date of Manufacture Waste of Electrical and Electronic Equipment REF Catalog Number SN Serial Number Accompanying Documents must be consulted Fragile Contents in Shipping Co...

Page 7: ...ronment Please retain this guide for future reference and to share with other users For additional copies of this manual or questions related to the Phoroptor contact your local authorized Reichert de...

Page 8: ...rom the Phoroptor Refer to Figure IS 4 8 Place the packaging in a safe place so that if transportation is required in the future it will be available Your new Phoroptor refracting instrument and acces...

Page 9: ...Lever 8 Auxiliary Lens Scale 9 Auxiliary Lens Knob 10 Corneal Aligning Device 11 Rotary Prism Unit 12 Cylinder Axis Knob 13 Cylinder Power Knob 14 Cross Cylinder Unit 15 Cylinder Power Scale 16 Cylin...

Page 10: ...ev L Instrument Setup continued Parts Identification continued 1 Forehead Rest 2 Spring Clip 2 1 Phoroptor Back Note The configuration of the Reichert Phoroptor is a registered trademark in the United...

Page 11: ...als Lighting is provided by energy efficient cool running LEDs light emitting diodes that require no maintenance Operators simply plug the Illuminated Phoroptor into the appropriate power receptacle t...

Page 12: ...that comes with the stand on the Phoroptor Arm Note The instrument cannot slide off the arm at this point but can be tilted forward or back 3 Tighten the Tilt Clamp Knob and the Phoroptor will be hel...

Page 13: ...e Shields Mounting Rotochart 1 To attach the Rotochart insert one pin into the hole slide the edge under the clip and insert the other pin into the second hole Refer to Figure SU 3 Note The Card Holde...

Page 14: ...imeters and diopters Leveling Phoroptor 1 Move the Phoroptor into a working position 2 Use the Leveling Adjustment Knob to adjust the horizontal position of the Phoroptor Turn the Leveling Knob until...

Page 15: ...n or occlude the apertures Note The right side of the Phoroptor the patient s left eye has white Maddox lenses and the green lens for the red green test The left side of the Phoroptor the patient s ri...

Page 16: ...line of the test chart should be level with or slightly below the lens apertures 6 With the above accomplished lock arm motion 7 Adjust Phoroptor Forehead Rest so that the patient s corneas are approx...

Page 17: ...be placed at a distance of 13 75 mm if the Phoroptor reading is to be directly applied to spectacle lens power 8 In the Phoroptor this condition is obtained when the zero setting of the sight in the...

Page 18: ...between 11 00 and 12 00 the correction factor according to the table is 0 62 Therefore the power of the correction lens is obtained by adding 0 62 to 11 50 which equals 10 88 diopters when the specta...

Page 19: ...48 57 11 00 12 25 37 51 64 78 11 00 12 24 36 46 57 68 12 00 15 30 45 61 77 93 12 00 14 28 42 55 67 80 13 00 18 35 53 72 91 1 10 13 00 16 33 48 64 78 94 14 00 21 41 62 84 1 06 1 29 14 00 19 38 55 74 9...

Page 20: ...copy subjective phorometry are generally made with the Vergence Levers in the extreme outward position Refer to Figure TM 1 2 At this setting the lens systems are parallel For shorter test distances t...

Page 21: ...xtreme outward position to the extreme inward position converges the instrument apertures for the near test at 16 inches At the same time aperture separation is decreased by 4mm 3 For PD settings grea...

Page 22: ...3 Rotation of the Weak Sphere Dial upward introduces more minus power or less plus Refer to Figure TM 3 Note Plus powers are indicated by black numbers minus powers by red numbers 4 An automatic pick...

Page 23: ...scale A quicker way rotate the Strong Sphere Control nasally one index position to introduce a value of 3 00 Rotate Weak Sphere Dial one index upward to reduce power to 2 75D To obtain a power of 7 00...

Page 24: ...M 5 4 A pair of 0 12D cylinders in accessory cells permits refinement to 1 8th D steps 5 A pair of 2 00D cylinders in accessory cells extends cylinder power range to 8 00D 6 Large 360 protractors arou...

Page 25: ...nce tests WMH or RMH Maddox Rod Horizontal White left eye red right eye For muscle balance tests RL Red Lens For binocular vision tests GL Green Lens For binocular vision tests O Open Aperture Second...

Page 26: ...in or base out prism Refer to Figure IU 6 5 The Finger Roll Knob rotates the prism lenses inside the Rotary Prism Unit and varies the magnitude of prism Refer to Figure IU 7 When the Finger Roll Knob...

Page 27: ...ubstituted 2 The power of the cross cylinder is engraved on the cell 3 Red dots indicate the minus axis and white dots the plus axis 4 A thumb operated Roll Knob provides for rapid flipping of cross c...

Page 28: ...the axis check Procedure 1 With the tentative sphere and correcting cylinder determined by retinoscopy and or the astig matic chart in place the Cross Cylinder Unit is positioned before the aperture o...

Page 29: ...k for axis following any modi fication made in the correcting cylinder axis and by following procedure as in a until final end point is reached c End point is reached i e correcting cylin der axis is...

Page 30: ...s perpendicular to correcting minus cylinder axis power of correcting cylinder is reduced c End point is obtained i e correcting cylin der power is correct when vision is equally impaired by flipping...

Page 31: ...TACTED WILL BE SOILED Sanitary Face Shields The semi permanent Face Shields furnished with the Phoroptor are made of white nylon and have a protective lens on them to help keep the inside of the Phoro...

Page 32: ...cleaning solution is desirable 4 Indirect illumination from a standard lamp bulb through the instrument apertures will help identify soiled lenses Lenses may be cleaned with the Phoroptor in place on...

Page 33: ...retaining washers 10 To lock new cell into position turn the screw 1 4 turn rotate washer so that flat side is 900 away from cell tighten the screw firmly 11 Repeat for the other screw and washer Repl...

Page 34: ...and MM 2 1 Using a 0 060 6 Spline Screwdriver remove the two spline screws 2 Lift off the retaining ring 3 Lift out the complete Cross Cylinder Lens Assembly 4 Reverse the procedure to install a new...

Page 35: ...pressure of the wheel in the dial index deter mines to a considerable degree the feel of the dial rotation and indexing Each may be adjusted from the outside of the in strument with no dismantling re...

Page 36: ...ct input voltage Light may be turned all the way down Turn the Rheostat Control knob all the way up Weak Sphere Dial Too Loose or Tight Tighten or loosen the coordinating screw on the bottom of the ph...

Page 37: ...Reading Distance 5 to 28 Centimeter and diopter scales included Auxiliary Dial 10 lenses plus 2 open apertures Electrical Illuminated Phoroptor Only Power Adaptor 11636 401 Power Adaptor 11636 405 In...

Page 38: ...dust and other contami nants do not get inside the unit Over time the presence of dust and other contaminants inside the unit may affect operation of the Phoroptor Disposal For proper disposal or recy...

Page 39: ...phthalmic Instrument Dealer Furthermore Reichert shall not be liable for costs of procurement of substitute goods by anyone How to File a Claim Any claim under this warranty must be made within 30 day...

Page 40: ...716 686 4555 Email reichert information ametek com www reichert com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Email info reichert de ametek c...

Reviews: