29
1400500-101 Rev. A
Instructions for Use
(continued)
DENSITY4/5 Remote Mode
(continued)
As long as the DENSITY4/5 is in remote mode, you can not see nor use the buttons
and
. You
can still turn on or off the light of the visual sample control by pressing
. To end the remote connection,
press the button local mode on the DENSITY4/5 display.
Figure IU-15, EDM in Remote Mode
Now you have the following options to control your DENSITY4/5 by using the Universal Display:
You can light the tube compartment with the button Light on. To activate the pump, press the Pump On
button.
If you click on the DENSITY4/5 temp. button you can adjust your wanted measuring temperature as well
as the allowed temperature deviation. Enter your wanted measuring temperature in the upper field and a
deviation in the lower field. This allowed Temperature deviance defines a range, the actual temperature
has to be in before a measurement starts. If the deviance is for example 0.5 °C and your Temperature is
set to 20.000 °C, the DENSITY4/5 won’t start measuring until the actual temperature reaches 20.500 °C.
Choose a deviation by pressing one of the according buttons.
Figure IU-16, Set DENSITY4/5 Temperature
-continued-