background image

10

1400500-101 Rev. A

Introduction

 (Continued)

Definition of Density 

(continued)

“Specific Gravity” (SG) (or “Relative Density” (RD)) is expressed in multiples of the density of some other 

standard material, has dimensionless quantity, and usually refers to water or air.

SG=   

P sample

P reference

__________

Oscillating U-Tube

The DENSITY4/5 has a built-in reference oscillating U-Tube, which is a new method of measuring density.  

Approximately 1.0 ml of sample is needed for a measurement and the device offers very high accuracy 

and resolution. Working with an oscillating U-Tube is ultra fast compared to other methods and true density 

is provided in the measurement. the system is fully automatic (calculations not necessary), so it is less 

vulnerable to human error.

The oscillating U-tube samples the viscosity of a fluid by vibrating a glass tube at a frequency which will 

measure the resistance of a fluid that is being deformed by either shear stress or tensional stress.  This 

sometimes is referred to as the “thickness” of a fluid and it has the following parameters:

• 

High viscosity → density over-reading

• 

Low viscosity → correct result

Another parameter to consider in the measurement process is the temperature of the fluid.  The tem-

perature of the fluid is very important. The built-in high accuracy temperature sensor guarantees precise 

measurement utilizing a Peltier-Element (temperature range 5°C to 95°C) to control the fluid temperature.

Some features of the temperature control system are:

•  Eliminates temperature drift.  After modifying the measuring temperature, the instrument 

remains stable as soon as the new temperature is reached.

• 

Simplifies measurement at different measuring temperatures.  

Using the temperature con-

trol at 20 °C provides consistent data related to one temperature.

(additional adjustment at 40 °C and 60 °C can be performed.  for example, once a year to re-adjust tem-

perature characteristics of the DENSITY4/5 sensor).

Figure IS-01,  Oscillating U-Tube Detail

Summary of Contents for DENSITY4

Page 1: ...REICHERT DENSITY4 5 User s Guide...

Page 2: ...ut notice Reichert Inc reserves the right to make changes in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 9001 Certified...

Page 3: ...se 18 Starting The DENSITY4 5 18 Keyboards 19 Icons 20 Menu Icons 20 Status Icons 20 Main Menu 21 Live Data 21 Last Measurement 21 Configuration 21 User Interface 22 Screen 22 Touch Calibration 22 Sou...

Page 4: ...NJURY OR DEATH WARNING THE OPERATOR OF THIS INSTRUMENT SHOULD USE IT IN STRICT ACCORDANCE WITH THE INSTRUCTIONS OUTLINED IN THIS USER S GUIDE REICHERT TECHNOLOGIES CANNOT GUARANTEE THE SAFETY OF THE O...

Page 5: ...VENTS OR STRONG CLEANING SOLUTIONS ON ANY PART OF THIS INSTRUMENT TO ENSURE THAT YOU DO NOT DAMAGE THE UNIT SEE THE MAINTENANCE SECTION OF THIS GUIDE FOR DETAILED CLEANING INSTRUCTIONS CAUTION DO NOT...

Page 6: ...YOU CAN ONLY TRANSFER DATA FROM THE DENSITY4 5 REFRACTOMETER TO YOUR COMPUTER BY INTERFACING THESE TWO DEVICES WITH THE SERIAL PORTS CAUTION DON T UNPLUG THE SERIAL CABLE WHILE DATA TRANSFER THIS COU...

Page 7: ...Guide Warning Alternating Current Power Manufacturer 2016 Date of Manufacture REF Catalog Number SN Serial Number Waste of Electrical and Electronic Equipment Compliance to Low Voltage Directive 2006...

Page 8: ...w the instructions in the guide to ensure optimum performance of your new instrument Please retain this guide for future reference and to share with other users For additional copies of this manual or...

Page 9: ...nging applications In addition this instrument has a built in moisture sensor an air pressure sensor and a very accurate temperature control system for extended usage in diffi cult environmental condi...

Page 10: ...e of a fluid that is being deformed by either shear stress or tensional stress This sometimes is referred to as the thickness of a fluid and it has the following parameters High viscosity density over...

Page 11: ...11 1400500 101 Rev A Instrument Setup Parts ID...

Page 12: ...er Cord for Measurement Head 5 Air Cartridge 6 Clips for Air Cartridge 7 130 ml Glass Cylinder 8 Package of 10 ml Syringes 9 Package of 5 ml Syringes 10 Connecting Hoses and Fittings 11 Air Filter 1 3...

Page 13: ...toggle switch Power Button to the ON position Connecting Air Cartridge Use the screws provided and align and attach the two metal clips provided onto the back side of the DEN SITY4 5 to accept the Air...

Page 14: ...Hose PTFE into the other inlet NOTE Always use the same inlet for waste and the same inlet for samples to avoid contamination Figure SU 03 Connection Sample Supply Connect the Connection Hose Density...

Page 15: ...above and flush it with distilled water After flushing it with water blow filtered air into the tubing that goes to the sample compartment to dry the compartment and remove any contaminants Connectin...

Page 16: ...ose Compressed Air and the Luer Adapter female as indicated below Figure SU 09 Clean with Dry Air Ensure that the open end of the hose from the outlet of the sample compartment is put into the Glass C...

Page 17: ...ort Connecting DENSITY4 5 and PC Connect the delivered Reichert Serial Connector Cable with the Reichert serial connector at the rear side of the device Note The Universal Display can also be connecte...

Page 18: ...the main screen Figure IU 01 Main Screen By pressing the menu button in the upper right corner you are reaching the Main Menu You have two options to login to your system 1 You can use a login pattern...

Page 19: ...efault keyboard with capital letters By pressing the abc button the keyboard changes to small letters If you press the 123 button you are reaching a keyboard with numbers and some of the most importan...

Page 20: ...ed In Not logged in as user Note You can click the menu icons to open or leave menus Status Icons The status icons showing you important information about processes on your device Remote Ongoing remot...

Page 21: ...nfiguration you can configure and adjust settings like date or time of your DENSITY4 5 In addition you can edit user settings to create new users and edit their permissions It is also possible to conf...

Page 22: ...calibration process Don t touch the screen during this time 3 Press the three glowing touch points that appear on the screen one at a time Figure IU 06 Touch Screen Calibration 1 4 Verify the calibra...

Page 23: ...ing the process dry it with the pump and check if it is clean inside Fill the tube with double distilled water or dry air now i Water fill in double distilled water with a syringe at the sample supply...

Page 24: ...e of your sample is displayed default unit is C If the temperature isn t stable its displayed in grey color If it reaches the adjusted temperature range the values are shown in black color In the thir...

Page 25: ...iation is for example 0 5 C and your target temperature is set to 20 000 C the DENSITY4 5 won t start measuring until the actual temperature reaches 20 500 C Choose a deviation by pressing one of the...

Page 26: ...e previous sample by rinsing the measuring cell with a sufficient amount 10 ml or more of the new sample Do not leave samples in the measuring cell longer than necessary Clean and dry the measuring ce...

Page 27: ...n sequential order do not skip a port Note The default connection for connecting only the DENSITY4 5 is COM1 If another instrument is con nected to the Serial port COM1 then any other sequential seria...

Page 28: ...splay Set the same COM port for communication with windows and mark your DENSITY4 5 as active Note The default connection for connecting only the DENSITY4 5 is COM1 If another instrument is con nected...

Page 29: ...nt with the button Light on To activate the pump press the Pump On button If you click on the DENSITY4 5 temp button you can adjust your wanted measuring temperature as well as the allowed temperature...

Page 30: ...LARLY PROVIDED WITH UPDATES DO NOT DISCONNECT INPUT POWER TO THE DEVICE OR SET THE ON OFF BUTTON TO OFF DURING THE UPDATE PROCESS OR DAMAGE TO THE UNIT OR SOFTWARE MAY OCCUR To update the software on...

Page 31: ...purifiers must take place after the measurement sequence Otherwise if you work with different samples with different chemical properties and these samples are not mixable and or can not be easily dis...

Page 32: ...he cleaning action of the water but this is not required in any case If you use a lab cleaner always follow the instructions given by the manufacturer Apply strongly alkaline lab cleaners pH above 10...

Page 33: ...es 2x Micro Fuse 3 15A Battery Interfaces Factory Connector includes Serial Communication RS232 adapter cable 1xEthernet LAN Standards Details on Request Dimensions 320 mm x 235 mm x 218 mm L x W x H...

Page 34: ...or handling of product PATENT WARRANTY If notified promptly in writing of any action brought against the purchaser based on a claim that the instru ment infringes a U S patent Reichert will defend suc...

Page 35: ...35 1400500 101 Rev A Notes...

Page 36: ...4555 Email reichert information ametek com www reichert com Authorized European Representative AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Email...

Reviews: