22
1400500-101 Rev. A
Instructions for Use
(continued)
Main Menu (
Continued
)
User Interface (
Continued
)
Touch Calibration
CAUTION: DO NOT DISCONNECT INPUT POWER OR SET THE DENSITY4/5 TO OFF DURING THE
CALIBRATION OR DAMAGE TO THE INSTRUMENT OR SOFTWARE MAY OCCUR.
Sometimes you need to (re-)calibrate the touchscreen of your DENSITY4/5. Follow the steps be-
low to calibrate:
1. Enter the menu Touchscreen calibration in the config menu. Read the shown message
carefully and begin the calibration by pressing start.
2. Pay attention to the countdown shown on the screen and wait three seconds for the device to
prepare the calibration process. Don’t touch the screen during this time!
3. Press the three glowing touch-points that appear on the screen one at a time.
Figure IU-06, Touch Screen Calibration (1)
4. Verify the calibration by pressing the next three points that are shown on the screen. Hold each
dot for about one second to verify calibration and to avoid a bad calibration. Time limit for each
calibration point is 60 seconds, after this time runs out, the calibration will restart.
If the verification process works well confirm the process by pressing OK.
Sound
Adjust the volume of the click sound of the device.
Calibration
Overview of the calibration data including current calibration factors that were created after zeroing.
You can start zeroing your device from this menu, the initial factory calibration never gets lost.