13
V_Encode no change
[Causes]
No signal response from the Encoder.
[Solutions]
(1) Replace the cable connected to the Encoder.
(2) Replace the motor module.
14
V_Position Over Pulse
[Causes]
The position command coordinate greatly differ from the Encoder
coordinate.
[Solutions]
(1) Check whether the terminal of the PLC Y00 Cable has dropped off.
(2) Replace the cable connected to the power of the motor.
(3) Replace the main board.
(4) Replace the motor module.
15
V_Home Sensor Fail
[Causes]
When performing the motor initial, the Home sensor has no response
and the motor collides with hard stop.
[Solutions]
(1) Check whether the cable has dropped off.
(2) Replace the sensor.
16
X_S
[Causes]
Manual X JOG+ operation exceeds the distance of the software.
[Solutions]
(1) Alarm Reset
(2) Manual X JOG- operation.
17
X_SoftLimt-
[Causes]
Manual X JOG- operation exceeds the distance of the
software.
[Solutions]
(1) Alarm Reset
(2) Manual X JOG+ operation.
Summary of Contents for MagCore Plus II
Page 1: ......
Page 5: ...Specification Operating Parameters Operating Environment Applications ...
Page 6: ...IQ OQ PQDocuments IQDocuments ...
Page 7: ... OQ Documents ...
Page 8: ... PQ Documents ...
Page 9: ......
Page 10: ...Accessories Installation 1 2 3 4 5 7 6 8 9 10 11 BeforeInstallation ...
Page 11: ... System Overview ...
Page 14: ... Installation ...
Page 15: ......
Page 33: ... Replace Fuse Please use the specified type of fuse tominimizetheriskofburning ...
Page 35: ......
Page 36: ... Disassemblethehousing 1 Loosen the two bottom screws to about 3mm ...
Page 37: ...2 Remove the remaining eight screws and take off the lid ...
Page 60: ...Electrical controlandengineeringsoftware ...
Page 61: ... Main board configuration ...
Page 66: ...10 Go to System Status and check the PLC version ...
Page 69: ...5 Go to System Status and check the HMI version ...
Page 78: ......
Page 79: ......
Page 80: ......