━ Heater Test
━ UV Sterilization
Please Select Heater Test Button.
Please put the thermometer into the Heat Block
Well 1 and press Next.
Select Test Temperature
Now the heater will heat up to 65° C or 90° C,
please check whether the temperature shown on
the thermometer is consistent with what's shown
on the screen.
The acceptable temperature range is 65± 5° C or
90± 5° C.
Please Select the Sterilization Button.
Please select the sterilization time and press Start
button.
Warning!
The UV lamp is dangerous. Please do not
look straight at the inner of the machine
when it is operating.
Please stop operation before material
change.
Please do not open this cover by
untrained operators.
Please be cautious when UV light is on.
Summary of Contents for MagCore Plus II
Page 1: ......
Page 5: ...Specification Operating Parameters Operating Environment Applications ...
Page 6: ...IQ OQ PQDocuments IQDocuments ...
Page 7: ... OQ Documents ...
Page 8: ... PQ Documents ...
Page 9: ......
Page 10: ...Accessories Installation 1 2 3 4 5 7 6 8 9 10 11 BeforeInstallation ...
Page 11: ... System Overview ...
Page 14: ... Installation ...
Page 15: ......
Page 33: ... Replace Fuse Please use the specified type of fuse tominimizetheriskofburning ...
Page 35: ......
Page 36: ... Disassemblethehousing 1 Loosen the two bottom screws to about 3mm ...
Page 37: ...2 Remove the remaining eight screws and take off the lid ...
Page 60: ...Electrical controlandengineeringsoftware ...
Page 61: ... Main board configuration ...
Page 66: ...10 Go to System Status and check the PLC version ...
Page 69: ...5 Go to System Status and check the HMI version ...
Page 78: ......
Page 79: ......
Page 80: ......