━ Replace O-Ring
━ Piston Test
Easy Maintenance
Easy Maintenance includes the maintenance items
which need to be done by users. These items include
sterilization, clearing, simple test and error message
report, etc. Users must perform these checkups
periodically to ensure the system status is always
good.
Please Select Replace O-Ring button.
1. Please disassemble the damaged O-Ring.
2. Replace a new O-Ring.
3. Greasing the O-Ring, and perform the Piston
Test.
The Piston test includes the tests for three Tip
types, which mainly test the tightness of Piston
O-Ring and Tip, and the smooth of taking the
Piston from the T-Rack.
Please Select Piston Test Button.
Maintenance:
Please perform the Piston Test and grease the
O-rings once a month.
Summary of Contents for MagCore Plus II
Page 1: ......
Page 5: ...Specification Operating Parameters Operating Environment Applications ...
Page 6: ...IQ OQ PQDocuments IQDocuments ...
Page 7: ... OQ Documents ...
Page 8: ... PQ Documents ...
Page 9: ......
Page 10: ...Accessories Installation 1 2 3 4 5 7 6 8 9 10 11 BeforeInstallation ...
Page 11: ... System Overview ...
Page 14: ... Installation ...
Page 15: ......
Page 33: ... Replace Fuse Please use the specified type of fuse tominimizetheriskofburning ...
Page 35: ......
Page 36: ... Disassemblethehousing 1 Loosen the two bottom screws to about 3mm ...
Page 37: ...2 Remove the remaining eight screws and take off the lid ...
Page 60: ...Electrical controlandengineeringsoftware ...
Page 61: ... Main board configuration ...
Page 66: ...10 Go to System Status and check the PLC version ...
Page 69: ...5 Go to System Status and check the HMI version ...
Page 78: ......
Page 79: ......
Page 80: ......