Content
Causes and solutions
4
DC Motor Up Time Out
[Causes]
The motor is not functional or the block plate of the sensor is at wrong
position.
[Solutions]
(1) Check whether the motor is functional.
(2) Check whether the cable dropped off.
(3) Adjust the position of the block plate of the sensor.
5
DC Motor Down Time Out
[Causes]
The motor is not functional or the block plate of the sensor is at wrong
position.
[Solutions]
(1) Check whether the motor is functional.
(2) Check whether the cable has dropped off.
(3) Adjust the position of the block plate of the sensor.
6
Y_S
[Causes]
Manual Y JOG+ operation exceeds the distance of the software.
[Solutions]
(1) Alarm Reset
(2) Manual Y JOG- operation.
7
Y_SoftLimt-
[Causes]
Manual Y JOG- operation exceeds the distance of the software.
[Solutions]
(1) Alarm Reset
(2) Manual Y JOG- operation.
━ Trouble Shooting for Alarm Code
Summary of Contents for MagCore Plus II
Page 1: ......
Page 5: ...Specification Operating Parameters Operating Environment Applications ...
Page 6: ...IQ OQ PQDocuments IQDocuments ...
Page 7: ... OQ Documents ...
Page 8: ... PQ Documents ...
Page 9: ......
Page 10: ...Accessories Installation 1 2 3 4 5 7 6 8 9 10 11 BeforeInstallation ...
Page 11: ... System Overview ...
Page 14: ... Installation ...
Page 15: ......
Page 33: ... Replace Fuse Please use the specified type of fuse tominimizetheriskofburning ...
Page 35: ......
Page 36: ... Disassemblethehousing 1 Loosen the two bottom screws to about 3mm ...
Page 37: ...2 Remove the remaining eight screws and take off the lid ...
Page 60: ...Electrical controlandengineeringsoftware ...
Page 61: ... Main board configuration ...
Page 66: ...10 Go to System Status and check the PLC version ...
Page 69: ...5 Go to System Status and check the HMI version ...
Page 78: ......
Page 79: ......
Page 80: ......