GP Monitor
34
Rev. F 10/2011
GP Series
3.4.12
Factory Defaults
If you need to reset the GP monitor to its factory default settings, you
can do so by pressing and holding the
button and then pressing
the button 3 times.
Parameter
Value
Setting Range
Remarks
Emissivity
0.95
0.1 - 1.09
For 0 - 5 V, 4 - 20 mA and
thermocouple inputs,
emissivity is not functional
Sensing Head/Device
type
H001
H001 - H007
Display Resolution
0
0 - 4
Setpoints (SP1, SP2)
off
According to Sensing
Head/Device type
see section 3.4.4, page
23
Peak hold
0
0-
998 s (999 s = ∞)
∞: reset by hardware
Valley hold
0
0-
998 s (999 s = ∞)
∞: reset by hardware
Averaging
0
0
– 60 s
Ambient Temperature
Compensation (t-amb)
0
-18 to 1200°C
(0 to 2200°F)
Offset
0
-50 to 50°C
(-99 to 99°F)
Table 10: Factory-set Default Values
3.5
Quick Reference
On the following pages are quick reference charts for setting up the
GP Monitor with sensing head and input devices and for adjusting
the monitor’s normal and advanced functions.