GP Monitor
GP Series
Rev. F 10/2011
5
3
GP Monitor
3.1
Technical Data
3.1.1
Measurement Specifications
Input Accuracy:
0 – 5 V input resolution: 1 mV,
accuracy: ± 2 mV
4 – 20 mA input resolution: 0.01 mA
accuracy: ± 0.02 mA
J, K, E, N, T thermocouple input:
± 0.5 % or ± 2°C, whichever is greater
R, S thermocouple input:
± 0.5 % or ± 3°C, whichever is greater
Output Accuracy:
4 – 20 mA output resolution: 0.014 mA
accuracy: ± 0.02 mA
J, K, E, N, T thermocouple output:
± 0.5 % or ± 2°C, whichever is greater
R, S thermocouple output:
± 0.5 % or ± 4°C, whichever is greater
Response Time:
300 ms (GPR/GPS system)
200 ms (all other)
Display:
4 digit LED
Emissivity:
0.10 to 1.09, adjustable in 0.01 increments –
H001 (GP heads) and H007 (XR heads) only
Warm-up Time:
5 s
Peak Hold:
0 to 998 s (999=infinite hold)
Valley Hold:
0 to 998 s (999=infinite hold)
Averaging:
0 to 60 s
Fail-Safe:
Full or low scale, depending upon system
failure, see section 6.1
Fail-Safe Operation
,