User Guide of RFL-C4000XZ~12000XZ
11
Figure 9 Output cable placed in the box
1) Place the output optical cable in a clean workbench;
Note: The following operations need to be carried out in a clean workbench. If the operation is
not carried out in a clean environment, it will cause great risk to the QP output optical cable;
operators need to wear finger cots to operate.
Figure 10 The output optical cable placed in the clean bench with the ventilation turned on
2) Remove the white dust cap and place it face down on new lint-free paper.
Note: The dust-free paper is for one-time use, and there is a risk of dust falling if it is used
multiple times.