Please note that in case of a failure of ART or of one of the ICPs, no CPP can be calculated any
more. In this situation no additional CPP alarm is activated in addition to the ART or ICP alarm,
and the CPP is no longer displayed.
Limiting values for ICP.
Alarm active / not active for ICP.
Limiting values for ICPA.
Alarm active / not active for ICPA.
Limiting values for CPP.
Alarm active / not active for CPP.
The settings rendered above are offered as default settings. For data on the basic settings of the alarm
limits, the alarm activation and the settings of the alarm limits, please refer to chapter 4.9.2 Factory settings.
In order to change the settings, use the arrow keys to navigate to the requested parameter and then press
the Menu key.