4 Operation of the MPR2
4.1 General aspects
4.1.1 Switching on the apparatus
Familiarise yourself with the instructions for use prior to using the MPR2.
If you have already used an MPR2, please observe the differences in the behaviour of the
appliance, in particular when setting the alarm limiting values. The behaviour of the appliance
depends on the revision status of the firmware. Please read chapter 4.6.2 Device diagnostics.
The apparatus may be operated only with the battery compartment closed.
Listen for the short peep when the apparatus is switched on. If this peep is not sounded, the
alarm system can be tested once again in the apparatus diagnosis mode. If no sound is produced
in this case either, monitoring is no longer possible with this apparatus.
After having switched on the apparatus, please check whether the date and time show correct
values. If this is not the case, please read chapter 4.6.5 Setting date and time.
If in the graphics or trend screen a graph passes along the upper or lower margin (measuring
range limit) of the diagram, the measured value may also be larger or smaller than shown
(overmodulation). This applies in particular, if a fixed scale is used instead of automatic scaling.
Prior to using the apparatus check the correct condition of the device and of the accessories. The apparatus
and the accessories must not be damaged or soiled. Please check for the presence of loose parts by
shaking or tipping the appliance, for example. If loose parts are present, the appliance must not be used. If
the MPR2 is operated on battery pack, please check the charge condition of the battery pack when
switching the apparatus on. The battery symbol
on the top right-hand side of the display can be
used to read the charge condition of the battery pack. A full bar indicates a charged battery pack. Given a
completely charged battery pack, the MPR2 can be operated up to two hours without a mains connection
(e.g. during transport of the patient). In this case, the green power On LED starts flashing (top left-hand
side on the operating panel).
If the MPR2 is to be operated by means of the mains adapter supplied, it has to be connected to the mains
voltage. Now connect the red connector of the mains adapter cable with the red port of the MPR2. Please
keep in mind that only one approved electrical appliance may be connected to the mains adapter.
Use key [5] to switch on the MPR2. After the apparatus has been switched on, it carries out a self test. This
key is also used to switch the appliance off. In order to prevent an unintentional shut-down, the key has to
be pressed for two seconds.
4.1.2 Screen selection
After the MPR2 has been switched on the start screen will appear; for the instructions to carry out the
settings, please refer to chapter 4.2 Start screen.
In the start screen, the settings are made such as patient documentation, how to select the memory mode,
et cetera. After the settings have been made, leave the start screen and skip to the overview screen which
is marked with the symbol in the information field of the display.