Instructions for Use RAUMEDIC
MPR2 logO
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 42 of 99
4.4.2.2.2 Activating and deactivating the alarms
If you want to activate or deactivate an alarm, please proceed as follows:
Use the arrow keys to navigate
into the table field, the settings of
which you want to change.
Subsequently press the Menu key.
Use the arrow keys to select the
requested setting. Subsequently
press the OK key.
The new setting has been
accepted, but it is not active yet.
Press the OK key to leave the
screen. Only then will all
modifications be active.
The modified settings become active only when the window is closed by pressing OK and after having
returned to the overview. Press the ESC key to discard all modified settings and to return to the overview.
4.4.2.3 Combine analogue outputs
The MPR2 has two analogue outputs. The analogue outputs can be used to index two random pressure
signals to a third-party system. Pressure signals could be two random pressures, but also the partial oxygen
pressure. Please observe chapter 3.4 Analogue outputs. The screen displaying the settings options of the
limiting values can be used to activate the connection of the analogue outputs. Another possibility is in
window
, which is described in chapter 4.6.1 Combine analogue outputs.
Step 1:
Select the analogue output Out1
or Out2 which you want to use to
index the ICP1 pressure signal to
the third-party system.
Step 2:
In the example you have selected
the output Out1. Follow the
instructions on the screen.
Step 3:
The cable connection between the
pressure input of the third-party
system and the analogue input
has been established. Zero the
third-party system. You have 180
seconds time to press the next
button. In case of a timeout the
system changes to step 1.