4.4.3.4.3 Switching off pressure measurement ICP2
If, in conjunction with a multi-parameter catheter, a technical fault occurs in the pressure measurement
ICP2 or ICP monitoring is not required, the measurement function ICP2 can be switched off separately.
In the overview window press key
ICP2 to change to the window for
setting the alarm limits and then
navigate to "Use channel". Then
press the menu key.
In the menu select the off option.
Confirm the selection with OK. The
settings are adopted when OK is
pressed for the second time.
The overview window shows that
the ICP2 pressure measurement
has been switched off.
4.4.3.5 pO2 limiting alarm values
The right-hand column of the display holds the navigation field, in which you can use the soft keys to carry
out further adjustments or to call off information or help.
Additional settings of the partial oxygen pressure can be accessed by pressing the pO2 key. If you have
connected additional sensors, other soft key designations will be displayed with the defined measuring
locations.
Parameter pO2
Range with alarm limits
Alarm activation
Setup measuring intervall
Press the Menu key, if you
want to modify the
highlighted field
In this window you can use the arrow keys to select whether you want to set alarm limits and whether the
limiting value monitor of the selected measuring value is to be active or not.
Do not set the limiting values to extreme figures which could render the alarm system useless.