Instructions for Use RAUMEDIC
MPR2 logO
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 72 of 99
4.9.2 Factory settings
4.9.2.1 Factory settings and adjustment ranges for physiological alarms
The factory setting for the display language is English.
The table rendered below shows the factory settings for physiological alarms as well as their adjustment
ranges.
Parameter
Measu-
ring
location
Factory settings
Adjustment range
Lower alarm
limit
Upper alarm
limit
Alarm
activated
Alarm limits *1
lower upper
Adjustment
step size
Pressure
ART SYS
100
180
Yes
-20
400
5
Pressure
ART MAP
80
140
Yes
-20
400
5
Pressure
ART DIA
60
100
Yes
-20
400
5
Pressure
CVP
0
20
Yes
-20
400
1
Pressure
ICP1 ICP
0
30
Yes
-20
400
1
Pressure
ICP1 ICPA
0
20
No
0
400
1
Pressure
ICP2 ICP
0
30
Yes
-20
400
1
Pressure
ICP2 ICPA
0
20
No
0
400
1
Pressure
ICP1/2 CPP
50
100
Yes
0
400
1
Pressure
GP1 SYS
100
180
Yes
-20
400
5
Pressure
GP1 MAP
80
140
Yes
-20
400
5
Pressure
GP1 DIA
60
100
Yes
-20
400
5
Pressure
GP2 SYS
100
180
Yes
-20
400
5
Pressure
GP2 MAP
80
140
Yes
-20
400
5
Pressure
GP2 DIA
60
120
Yes
-20
400
5
Pressure
GP3 SYS
100
180
Yes
-20
400
5
Pressure
GP3 MAP
80
140
Yes
-20
400
5
Pressure
GP3 DIA
60
120
Yes
-20
400
5
Oxygen
pO2
10
80
Yes
0
200
1
Tempera-
ture
T1
35.8
37.5
Yes
15.0
45.0
0.1
Tempera-
ture
T2
35.8
37.5
Yes
15.0
45.0
0.1
Table 6: Factory settings
*1 The alarm limits can be adjusted freely within the range given; the minimum distance between upper
and lower limit has to be one step width.
4.9.2.2 Alarm signal inactivation
The global AUDIO PAUSING has been set to a duration of 10 minutes.