Remote control commands
R&S
®
RTO6
2435
User Manual 1801.6687.02 ─ 05
The effect depends on
TRIGger<m>:PARallel:WIDTh:RANGe
.
●
For the ranges SHORter and LONGer, the width defines the maximum and mini-
mum pulse width, respectively.
●
For the ranges WITHin and OUTSide, the width defines the center of a range
which is defined by the limits
TRIGger<m>:PARallel:WIDTh:DELTa
Parameters:
<Width>
Range:
200E-12 to 10000
Increment: 200E-9
*RST:
5E-9
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:PARallel:WIDTh:DELTa
<WidthDelta>
Defines a range around the given width value. the setting is relevant if
is set to WITHin or OUTSide. The width is set with
TRIGger<m>:PARallel:WIDTh:WIDTh
.
Parameters:
<WidthDelta>
Range:
0 to 432
Increment: 600E-12
*RST:
0
Default unit: s
Usage:
Asynchronous command
24.18.4.4
Timeout trigger
See also:
●
TRIGger<m>:PARallel:TIMeout:EXPRession[:DEFine]
TRIGger<m>:PARallel:TIMeout:RANGe
......................................................................... 2435
TRIGger<m>:PARallel:TIMeout:TIME
.............................................................................2436
TRIGger<m>:PARallel:TIMeout:RANGe
<TimeoutMode>
Sets the state condition.
Parameters:
<TimeoutMode>
HIGH | LOW | EITHer
HIGH
The level of a digital channel stays above the threshold, or the
logical expression for LOGic trigger source is true.
LOW
The level of a digital channel stays below the threshold, or the
logical expression for LOGic trigger source is false.
EITHer
The signal state remains unchanged.
Mixed signal option (MSO, R&S
RTO6-B1)