Remote control commands
R&S
®
RTO6
2436
User Manual 1801.6687.02 ─ 05
*RST:
HIGH
Usage:
Asynchronous command
TRIGger<m>:PARallel:TIMeout:TIME
<Time>
Defines the time limit for the timeout at which the instrument triggers.
Parameters:
<Time>
Range:
100E-12 to 10000
Increment: 100E-9
*RST:
100E-9
Default unit: s
Usage:
Asynchronous command
24.18.4.5
Setup & Hold trigger
See also:
●
TRIGger<m>:PARallel:SETHold:CSOurce[:VALue]
TRIGger<m>:PARallel:SETHold:CSOurce:EDGE
............................................................ 2436
TRIGger<m>:PARallel:SETHold:HTIMe
..........................................................................2436
TRIGger<m>:PARallel:SETHold:STIMe
.......................................................................... 2437
TRIGger<m>:PARallel:SETHold:CSOurce:EDGE
<ClockEdge>
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time reference point for the setup and hold time measurement.
Parameters:
<ClockEdge>
POSitive | NEGative | EITHer
*RST:
POSitive
Usage:
Asynchronous command
TRIGger<m>:PARallel:SETHold:HTIMe
<HoldTime>
Sets the minimum time
after
the clock edge while data should be stable and not
change its state.
See also:
Parameters:
<HoldTime>
Range:
-99.8E-9 to 100E-9
Increment: 1E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
Mixed signal option (MSO, R&S
RTO6-B1)